\ No newline at end of file +\chapter{Procedures} + +\section{Aligining TOPAS-C} + +\section{Aligning Spitfire PRO} + +\section{Air Handling} + +\section{Six Month Maintenance} + +\section{Tuning MicroHR Monochromator} + +Visible Grating. % + +Align the HeNe as perpendicular as possible to the monochromator entrance slit. % + +Move the grating angle until the HeNe falls on the exit slit. % + +Shine a flashlight through the entrance slit and observe the colour on the exit slit: if white, +then you are at 0-order (0 nm), if red, then you are at 1st order (632.8 nm). % + +Go to 0-order, narrow the slits, and slowly adjust the angle until the HeNe is going through the +exit slit. % + +Go to Jovin Yvon/utilities and find the motor configuration program. % + +In the Gratings tab, select the 1st grating (1200 line density) and hit Calibrate. % + +In theoretical wavelength, enter 0 nm. % + +In experimental wavelength, enter the wavelength you observe from the control program. % + +Hit set. % |