aboutsummaryrefslogtreecommitdiff
path: root/procedures/chapter.tex
diff options
context:
space:
mode:
Diffstat (limited to 'procedures/chapter.tex')
-rw-r--r--procedures/chapter.tex34
1 files changed, 33 insertions, 1 deletions
diff --git a/procedures/chapter.tex b/procedures/chapter.tex
index 3cb42bf..4ac67ed 100644
--- a/procedures/chapter.tex
+++ b/procedures/chapter.tex
@@ -1 +1,33 @@
-\chapter{Procedures} \ No newline at end of file
+\chapter{Procedures}
+
+\section{Aligining TOPAS-C}
+
+\section{Aligning Spitfire PRO}
+
+\section{Air Handling}
+
+\section{Six Month Maintenance}
+
+\section{Tuning MicroHR Monochromator}
+
+Visible Grating. %
+
+Align the HeNe as perpendicular as possible to the monochromator entrance slit. %
+
+Move the grating angle until the HeNe falls on the exit slit. %
+
+Shine a flashlight through the entrance slit and observe the colour on the exit slit: if white,
+then you are at 0-order (0 nm), if red, then you are at 1st order (632.8 nm). %
+
+Go to 0-order, narrow the slits, and slowly adjust the angle until the HeNe is going through the
+exit slit. %
+
+Go to Jovin Yvon/utilities and find the motor configuration program. %
+
+In the Gratings tab, select the 1st grating (1200 line density) and hit Calibrate. %
+
+In theoretical wavelength, enter 0 nm. %
+
+In experimental wavelength, enter the wavelength you observe from the control program. %
+
+Hit set. %