aboutsummaryrefslogtreecommitdiff
path: root/procedures/chapter.tex
diff options
context:
space:
mode:
Diffstat (limited to 'procedures/chapter.tex')
-rw-r--r--procedures/chapter.tex106
1 files changed, 74 insertions, 32 deletions
diff --git a/procedures/chapter.tex b/procedures/chapter.tex
index 9c9bd32..fcac88e 100644
--- a/procedures/chapter.tex
+++ b/procedures/chapter.tex
@@ -1,8 +1,43 @@
\chapter{Procedures}
+\clearpage
+\section{Starting the Millenia V CW LASER.}
+
+The Millennia has problems with ‘thermal runaway’ upon startup if a user naively sends the laser to
+4.0 W in power mode (thermal runaway causes the infamous FAULT 146 system shut off, power adjust
+timeout error). %
+The following procedure wakes the Millenia up gently enough to prevent thermal runaway. %
+It assumes that the Millenia starts completely shutdown and cold. %
+
+\begin{denumerate}
+ \item Check desiccant inside Millenia, replace if pink.
+ \item Ensure that Millenia shutter is closed such that no light is going downstream. %
+ \item Flip orange power switch on diode box.
+ \item Wait for system to warm up ($\sim$30 minutes).
+ \begin{denumerate}
+ \item ``System Warming Up'' message should appear on control box.
+ \item Warm-up is finished when percentage complete indicator reaches 100.
+ \item The Millenia will start in SP Current mode.
+ \end{denumerate}
+ \item Record diode hours (choose ``Info'', scroll down).
+ \item Turn Millennia to 1.0 W in ``power mode'', wait for equilibration.
+\end{denumerate}
+
+
+Turn Millennia to 1.0 W in “power mode”, wait for it to equilibrate. This entails the diode temperature and current being stable (within 0.1) over 5 minutes.
+Turn Millennia to “current mode”.
+Slowly ramp current until you have ~4 W (or more) (~70%)
+Allow the diode temperature to stabilize while remaining in “current mode” (check every 5 minutes until it doesn't change by more than 0.1)
+Turn Millennia to 4.0 W “power mode”.
+Wait 10 minutes for extra assurance of equilibration.
+Ensure again that temperature and current are not changing (over ~5 minute timescale).
+Measure and record actual Millenia output power, currents, temperatures.
+These data should be recorded in the GoogleSheet document for Millennia.
+
+\clearpage
\section{Calibrating the 407A}
-Calibrating the 407A
+Calibrating the 407.A
You may sometimes notice that the zero position changes dramatically from sensitivity to
sensitivity with the 407A. If this happens, iterate through the following until zero stays
@@ -13,28 +48,25 @@ Use the fine adjust (knob on side) to zero the 407A on the highest sensitivity
Use the front adjust (flathead screwdriver needed) to zero on the lowest sensitivity
\clearpage
-\section{Lytron Kodiak RC006}
+\section{Lytron Kodiak RC006
+We have one Lytron Kodiak RC006: Model Number RC006G03BB1C002, Serial Number 739383-02.
-We have one Lytron Kodiak RC006:
-Model Number RC006G03BB1C002
-Serial Number 739383-02
-
-Regular Chiller Maintenance:
+Regular chiller maintenance:
\begin{denumerate}
\item Gather supplies.
- \begin{enumerate}
+ \begin{denumerate}
\item 1 gallon distilled water (do not use deionized)
\item 1 gallon Nalco 460-PCCL104 (the pink stuff)
- \end{enumerate}
+ \end{denumerate}
\item Drain system completely.
- \begin{enumerate}
+ \begin{denumerate}
\item Turn off chiller.
\item Break tubing at push-to-connect, plugging return.
\item Allow chiller to run until liquid stops flowing. You will get a pressure error, ignore
(silence) it.
\item Use tube to mouth-syphon remaining liquid from within chiller
\item Reconnect at push-to-connect
- \end{enumerate}
+ \end{denumerate}
\item Fill chiller with distilled water (should require <1 gallon), let run for 30 minutes.
\item Drain system again.
\item Replace filter.
@@ -59,38 +91,48 @@ Grey: Serial Number 3E1161245
White and Blue: Serial Number 4K1050550
-Regular Chiller Maintenance:
-\begin{enumerate}
- \item Gather supplies
-\end{enumerate}
-
-maintenance directions
-Gather supplies
-Filter - sold in the stockroom
-1 gallon of Nalco 460-PCCL104 (the pink stuff)
-Drain system completely
-Disconnect red water line and allow chiller to pump water out.
-Push nipple of male end with flat object so check valve is released
-Take off of filter
-Push down to allow water out
-Check filters
-If air filter dirty, wash with water (let dry after washing)
-If water filter is dirty, replace
-Reassemble, fill with Nalco 460-PCCL104 (the pink stuff)
-Turn chiller on and top-off.
-
-\section{NesLab Merlin M33}
+Regular chiller maintenance:
+\begin{denumerate}
+ \item Gather supplies.
+ \begin{denumerate}
+ \item Filter (sold in stockroom).
+ \item 1 gallon Nalco 460-PCCL104 (the pink stuff).
+ \end{denumerate}
+ \item Drain system completely.
+ \begin{denumerate}
+ \item Disconnect red water line and allow chiller to pump water out.
+ \item Push nipple of male end with flat object to release check valve.
+ \end{denumerate}
+ \item Check filters.
+ \begin{denumerate}
+ \item If air filter dirty, wash with water (let dry after washing).
+ \item If water filter is dirty, replace.
+ \end{denumerate}
+ \item Reassemble.
+ \begin{denumerate}
+ \item Fill with Nalco 460-PCCL104 (the pink stuff).
+ \item Turn on, top-off with distilled water.
+ \end{denumerate}
+\end{denumerate}
+% TODO: figure
+\clearpage
+\section{NesLab Merlin M33}
+\clearpage
\section{Aligining TOPAS-C}
+\clearpage
\section{Aligning Spitfire PRO}
+\clearpage
\section{Air Handling}
+\clearpage
\section{Six Month Maintenance}
+\clearpage
\section{Tuning MicroHR Monochromator}
Visible Grating. %