1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
246
247
248
249
250
251
252
253
254
255
256
257
258
259
260
261
262
263
264
265
266
267
268
269
270
271
272
273
274
275
276
277
278
279
280
281
282
283
284
285
286
287
288
289
290
291
292
293
294
295
296
297
298
299
300
301
302
303
304
305
306
307
308
309
310
311
312
313
314
315
316
317
318
319
320
321
322
323
324
325
326
327
328
329
330
331
332
333
334
335
336
337
338
339
340
341
342
343
344
345
346
347
348
349
350
351
352
353
354
355
356
357
358
359
360
361
362
363
364
365
366
367
368
369
370
371
372
373
374
375
376
377
378
379
380
381
382
383
384
385
386
387
388
389
390
391
392
393
394
395
396
397
398
399
400
401
402
403
404
405
406
407
408
409
410
411
412
413
414
415
416
417
418
419
420
421
422
423
424
425
426
427
428
429
430
431
432
433
434
435
436
437
438
439
440
441
442
443
444
445
446
447
448
449
450
451
452
453
454
455
456
457
458
459
460
461
462
463
464
465
466
467
468
469
470
471
472
473
474
475
476
477
478
479
480
481
482
483
484
485
486
487
488
489
490
491
492
493
494
495
496
497
498
499
500
501
502
503
504
505
506
507
508
509
510
511
512
513
514
515
516
517
518
519
520
521
522
523
524
525
526
527
528
529
530
531
532
533
534
535
536
537
538
539
540
541
542
543
544
545
546
547
548
549
550
551
552
553
554
555
556
557
558
559
560
561
562
563
564
565
566
567
568
569
570
571
572
573
574
575
576
577
578
579
580
581
582
583
584
585
586
587
588
589
590
591
592
593
594
595
596
597
598
599
600
601
602
603
604
605
606
607
608
609
610
611
612
613
614
615
616
617
618
619
620
621
622
623
624
625
626
627
628
629
630
631
632
633
634
635
636
637
638
639
640
641
642
643
644
645
646
647
648
649
650
651
652
653
654
655
656
657
658
659
660
661
662
663
664
665
666
667
668
669
670
671
672
673
674
675
676
677
678
679
680
681
682
683
684
685
686
687
688
689
690
691
692
693
694
695
696
697
698
699
700
701
702
703
704
705
706
707
708
709
710
711
712
713
714
715
716
717
718
719
720
721
722
723
724
725
726
727
728
729
730
731
732
733
734
735
736
737
738
739
740
741
742
743
744
745
746
747
748
749
750
751
752
753
754
755
756
757
758
759
760
761
762
763
764
765
766
767
768
769
770
771
772
773
774
775
776
777
778
779
780
781
782
783
784
785
786
787
788
789
790
791
792
793
794
795
796
797
798
799
800
801
802
803
804
805
806
807
808
809
810
811
812
813
814
815
816
817
818
819
820
821
822
823
824
825
826
827
828
829
830
831
832
833
834
835
836
837
838
839
840
841
842
843
844
845
846
847
848
849
850
851
852
853
854
855
856
857
858
859
860
861
862
863
864
865
866
867
868
869
870
871
872
873
874
875
876
877
878
879
880
881
882
883
884
885
886
887
888
889
890
891
892
893
894
895
896
897
898
899
900
901
902
903
904
905
906
907
908
909
910
911
912
913
914
915
916
917
918
919
920
921
922
923
924
925
926
927
928
929
930
931
932
933
934
935
936
937
938
939
940
941
942
943
944
945
946
947
948
949
950
951
952
953
954
955
956
957
958
959
960
961
962
963
964
965
966
967
968
969
970
971
972
973
974
975
976
977
978
979
980
981
982
983
984
985
986
987
988
989
990
991
992
993
994
995
996
997
998
999
1000
1001
1002
1003
1004
1005
1006
1007
1008
1009
1010
1011
1012
1013
1014
1015
1016
1017
1018
1019
1020
1021
1022
1023
1024
1025
1026
1027
1028
1029
1030
1031
1032
1033
1034
1035
1036
1037
1038
1039
1040
1041
1042
1043
1044
1045
1046
1047
1048
1049
1050
1051
1052
1053
1054
1055
1056
1057
1058
1059
1060
1061
1062
1063
1064
1065
1066
1067
1068
1069
1070
1071
1072
1073
1074
1075
1076
1077
1078
1079
1080
1081
1082
1083
1084
1085
1086
1087
1088
1089
1090
1091
1092
1093
1094
1095
1096
1097
1098
1099
1100
1101
1102
1103
1104
1105
1106
1107
1108
1109
1110
1111
1112
1113
1114
1115
1116
1117
1118
1119
1120
1121
1122
1123
1124
1125
1126
1127
1128
1129
1130
1131
1132
1133
1134
1135
1136
1137
1138
1139
1140
1141
1142
1143
1144
1145
1146
1147
1148
1149
1150
1151
1152
1153
1154
1155
1156
1157
1158
1159
1160
1161
1162
1163
1164
1165
1166
1167
1168
1169
1170
1171
1172
1173
1174
1175
1176
1177
1178
1179
1180
1181
1182
1183
1184
1185
1186
1187
1188
1189
1190
1191
1192
1193
1194
1195
1196
1197
1198
1199
1200
1201
1202
1203
1204
1205
1206
1207
1208
1209
1210
1211
1212
1213
1214
1215
1216
1217
1218
1219
1220
1221
1222
1223
1224
1225
1226
1227
1228
1229
1230
1231
1232
1233
1234
1235
1236
1237
1238
1239
1240
1241
1242
1243
1244
1245
1246
1247
1248
1249
1250
1251
1252
1253
1254
1255
1256
1257
1258
1259
1260
1261
1262
1263
1264
1265
1266
1267
1268
1269
1270
1271
1272
1273
1274
1275
1276
1277
1278
1279
1280
1281
1282
1283
1284
1285
1286
1287
1288
1289
1290
1291
1292
1293
1294
1295
1296
1297
1298
1299
1300
1301
1302
1303
1304
1305
1306
1307
1308
1309
1310
1311
1312
1313
1314
1315
1316
1317
1318
1319
1320
1321
1322
1323
1324
1325
1326
1327
1328
1329
1330
1331
1332
1333
1334
1335
1336
1337
1338
1339
1340
1341
1342
1343
1344
1345
1346
1347
1348
1349
1350
1351
1352
1353
1354
1355
1356
1357
1358
1359
1360
1361
1362
1363
1364
1365
1366
1367
1368
|
\chapter{Disentangling material and instrument response} \label{cha:mix}
\textit{This Chapter borrows extensively from \textcite{KohlerDanielDavid2017a}.}
Ultrafast spectroscopy is often collected in the mixed frequency/time domain, where pulse durations
are similar to system dephasing times. %
In these experiments, expectations derived from the familiar driven and impulsive limits are not
valid. %
This work simulates the mixed-domain Four Wave Mixing response of a model system to develop
expectations for this more complex field-matter interaction. %
We explore frequency and delay axes. %
We show that these line shapes are exquisitely sensitive to excitation pulse widths and delays. %
Near pulse overlap, the excitation pulses induce correlations which resemble signatures of dynamic
inhomogeneity. %
We describe these line shapes using an intuitive picture that connects to familiar field-matter
expressions. %
We develop strategies for distinguishing pulse-induced correlations from true system
inhomogeneity. %
These simulations provide a foundation for interpretation of ultrafast experiments in the mixed
domain. %
\section{Introduction} % =========================================================================
Ultrafast spectroscopy is based on using nonlinear interactions, created by multiple ultrashort
($10^{-9}-10^{-15}$ s) pulses, to resolve spectral information on timescales as short as the pulses
themselves. \cite{RentzepisPM1970a, MukamelShaul2000a} %
The ultrafast specta can be collected in the time domain or the frequency
domain. \cite{ParkKisam1998a} %
Time-domain methods/ scan the pulse delays to resolve the free induction decay
(FID). \cite{GallagherSarahM1998a} %
The Fourier Transform of the FID gives the ultrafast spectrum. %
Ideally, these experiments are performed in the impulsive limit where FID dominates the
measurement. %
FID occurs at the frequency of the transition that has been excited by a well-defined, time-ordered
sequence of pulses. %
Time-domain methods are compromised when the dynamics occur on faster time scales than the
ultrafast excitation pulses. %
As the pulses temporally overlap, FID from other pulse time-orderings and emission driven by the
excitation pulses both become important. %
These factors are responsible for the complex ``coherent artifacts'' that are often ignored in
pump-probe and related methods. \cite{LebedevMV2007a, VardenyZ1981a, JoffreM1988a,
PollardW1992a} %
Dynamics faster than the pulse envelopes are best measured using line shapes in frequency domain
methods. %
Frequency-domain methods scan pulse frequencies to resolve the ultrafast spectrum
directly. \cite{DruetSAJ1979a, OudarJL1980a} %
Ideally, these experiments are performed in the driven limit where the steady state dominates the
measurement. %
In the driven limit, all time-orderings of the pulse interactions are equally important and FID
decay is negligible. %
The output signal is driven at the excitation pulse frequencies during the excitation pulse
width. %
Frequency-domain methods are compromised when the spectral line shape is narrower than the
frequency bandwidth of the excitation pulses. %
Dynamics that are slower than the pulse envelopes can be measured in the time domain by resolving
the phase oscillations of the output signal during the entire FID decay. %
There is also the hybrid mixed-time/frequency-domain approach, where pulse delays and pulse
frequencies are both scanned to measure the system response. %
This approach is uniquely suited for experiments where the dephasing time is comparable to the
pulse durations, so that neither frequency-domain nor time-domain approaches excel on their
own. \cite{OudarJL1980a, WrightJohnCurtis1997b, WrightJohnCurtis1991a} %
In this regime, both FID and driven processes are important. \cite{PakoulevAndreiV2006a} %
Their relative importance depends on pulse frequencies and delays. %
Extracting the correct spectrum from the measurement then requires a more complex analysis that
explicitly treats the excitation pulses and the different
time-orderings. \cite{PakoulevAndreiV2007a, KohlerDanielDavid2014a, GelinMaximF2009b} %
Despite these complications, mixed-domain methods have a practical advantage: the dual frequency-
and delay-scanning capabilities allow these methods to address a wide variety of dephasing
rates. %
The relative importance of FID and driven processes and the changing importance of different
coherence pathways are important factors for understanding spectral features in all ultrafast
methods. %
These methods include partially-coherent methods involving intermediate populations such as
pump-probe \cite{HammPeter2000a}, transient grating \cite{SalcedoJR1978a, FourkasJohnT1992a,
FourkasJohnT1992b}, transient absorption/reflection \cite{AubockGerald2012a, BakkerHJ2002a}, photon
echo \cite{DeBoeijWimP1996a, PattersonFG1984a, TokmakoffAndrei1995a}, two dimensional-infrared
spectroscopy (2D-IR) \cite{HammPeter1999a, AsplundMC2000a, ZanniMartinT2001a}, 2D-electronic
spectroscopy (2D-ES) \cite{HyblJohnD2001b, BrixnerTobias2004a}, and three pulse photon echo peak
shift (3PEPS) \cite{EmdeMichelF1998a, DeBoeijWimP1996a, DeBoeijWimP1995a, ChoMinhaeng1992a,
PassinoSeanA1997a} spectroscopies. %
These methods also include fully-coherent methods involving only coherences such as Stimulated
Raman Spectroscopy (SRS) \cite{YoonSangwoon2005a, McCamantDavidW2005a}, Doubly Vibrationally
Enhanced (DOVE) \cite{ZhaoWei1999a, ZhaoWei1999b, ZhaoWei2000a, MeyerKentA2003a,
DonaldsonPaulMurray2007b, DonaldsonPaulMurray2008a, FournierFrederic2008a},
Triply Resonant Sum Frequency (TRSF) \cite{BoyleErinSelene2013a, BoyleErinSelene2013b,
BoyleErinSelene2014a}, Sum Frequency Generation (SFG) \cite{LagutchevAlexi2007a}, Coherent
Anti-Stokes Raman Spectroscopy (CARS) \cite{CarlsonRogerJohn1990b, CarlsonRogerJohn1990c,
CarlsonRogerJohn1991a}, and other coherent Raman methods \cite{SteehlerJK1985a}. %
This paper focuses on understanding the nature of the spectral changes that occur in Coherent
Multidimensional Spectroscopy (CMDS) as experiments transition between the two limits of frequency-
and time-domain methods. %
CMDS is a family of spectroscopies that use multiple delay and/or frequency axes to extract
homogeneous and inhomogeneous broadening, as well as detailed information about spectral diffusion
and chemical changes. \cite{KwacKijeong2003a, WrightJohnCurtis2016a} %
For time-domain CMDS (2D-IR, 2D-ES), the complications that occur when the impulsive approximation
does not strictly hold has only recently been addressed. \cite{PerlikVaclav2017a,
SmallwoodChristopherL2016a} %
Frequency-domain CMDS methods, referred to herein as multi-resonant CMDS (MR-CMDS), have similar
capabilities for measuring homogeneous and inhomogeneous broadening. %
Although these experiments are typically described in the driven
limit, \cite{GallagherSarahM1998a, FourkasJohnT1992a, FourkasJohnT1992b} many of the experiments
involve pulse widths that are comparable to the widths of the system. \cite{MeyerKentA2003a,
DonaldsonPaulMurray2007b, PakoulevAndreiV2009a, ZhaoWei1999a, CzechKyleJonathan2015a,
KohlerDanielDavid2014a} %
MR-CMDS then becomes a mixed-domain experiment whereby resonances are characterized with marginal
resolution in both frequency and time. %
For example, DOVE spectroscopy involves three different pathways \cite{WrightJohnCurtis2003a} whose
relative importance depends on the relative importance of FID and driven
responses. \cite{DonaldsonPaulMurray2010a} %
In the driven limit, the DOVE line shape depends on the difference between the first two pulse
frequencies so the line shape has a diagonal character that mimics the effects of inhomogeneous
broadening. %
In the FID limit where the coherence frequencies are defined instead by the transition, the
diagonal character is lost. %
Understanding these effects is crucial for interpreting experiments, yet these effects have not
been characterized for MR-CMDS. %
This work considers the third-order MR-CMDS response of a 3-level model system using three
ultrafast excitation beams with the commonly used four-wave mixing (FWM) phase-matching condition,
$\vec{k}_\text{out} = \vec{k}_1 - \vec{k}_2 + \vec{k}_{2'}$. %
Here, the subscripts represent the excitation pulse frequencies, $\omega_1$ and $\omega_2 =
\omega_{2'}$. %
These experimental conditions were recently used to explore line shapes of excitonic
systems, \cite{KohlerDanielDavid2014a, CzechKyleJonathan2015a} and have been developed on
vibrational states as well. \cite{MeyerKentA2004a} %
Although MR-CMDS forms the context of this model, the treatment is quite general because the phase
matching condition can describe any of the spectroscopies mentioned above with the exception of SFG
and TRSF, for which the model can be easily extended. %
We numerically simulate the MR-CMDS response with pulse durations at, above, and below the system
coherence time. %
To highlight the role of pulse effects, we build an interpretation of the full MR-CMDS response by
first showing how finite pulses affect the evolution of a coherence, and then how finite pulses
affect an isolated third-order pathway. %
When considering the full MR-CMDS response, we show that spectral features change dramatically as a
function of delay, even for a homogeneous system with elementary dynamics. %
Importantly, the line shape can exhibit correlations that mimic inhomogeneity, and the temporal
evolution of this line shape can mimic spectral diffusion. %
We identify key signatures that can help differentiate true inhomogeneity and spectral diffusion
from these measurement artifacts. %
\section{Theory} % ===============================================================================
\begin{figure}
\includegraphics[width=0.5\linewidth]{"mixed_domain/WMELs"}
\caption[Sixteen triply-resonant Liouville pathways.]{
The sixteen triply-resonant Liouville pathways for the third-order response of the system used
here.
Time flows from left to right.
Each excitation is labeled by the pulse stimulating the transition; excitatons with $\omega_1$
are yellow, excitations with $\omega_2=\omega_{2'}$ are purple, and the final emission is gray.
}
\label{mix:fig:WMELs}
\end{figure}
We consider a simple three-level system (states $n=0,1,2$) that highlights the multidimensional
line shape changes resulting from choices of the relative dephasing and detuning of the system and
the temporal and spectral widths of the excitation pulses. %
For simplicity, we will ignore population relaxation effects: $\Gamma_{11}=\Gamma_{00}=0$. %
The electric field pulses, $\left\{E_l \right\}$, are given by:
\begin{equation} \label{mix:eqn:E_l}
E_l(t; \omega_l, \tau_l, \vec{k}_l \cdot z) = \frac{1}{2}\left[c_l(t-\tau_l)e^{i\vec{k}_l\cdot z}e^{-i\omega_l(t-\tau_l)} + c.c. \right],
\end{equation}
where $\omega_l$ is the field carrier frequency, $\vec{k}_l$ is the wavevector, $\tau_l$ is the
pulse delay, and $c_l$ is a slowly varying envelope. %
In this work, we assume normalized (real-valued) Gaussian envelopes: %
\begin{equation}
c_l(t) = \frac{1}{\Delta_t}\sqrt{\frac{2\ln 2}{2\pi}} \exp\left(-\ln 2 \left[\frac{t}{\Delta_t}\right]^2\right),
\end{equation}
where $\Delta_t$ is the temporal FWHM of the envelope intensity. %
We neglect non-linear phase effects such as chirp so the FWHM of the frequency bandwidth is
transform limited: $\Delta_{\omega}\Delta_t=4 \ln 2 \approx 2.77$, where $\Delta_{\omega}$ is the
spectral FWHM (intensity scale). %
The Liouville-von Neumann Equation propagates the density matrix, $\bm{\rho}$:
\begin{equation} \label{mix:eqn:LVN}
\frac{d\bm{\rho}}{dt} = -\frac{i}{\hbar}\left[\bm{H_0} + \bm{\mu}\cdot \sum_{l=1,2,2^\prime} E_l(t), \bm{{\rho}}\right] + \bm{\Gamma \rho}.
\end{equation}
Here $\bm{H_0}$ is the time-independent Hamiltonian, $\bm{\mu}$ is the dipole superoperator, and
$\bm{\Gamma}$ contains the pure dephasing rate of the system. %
We perform the standard perturbative expansion of \autoref{mix:eqn:LVN} to third order in the
electric field interaction \cite{MukamelShaul1995a, YeeTK1978a, OudarJL1980a, ArmstrongJA1962a,
SchweigertIgorV2008a} and restrict ourselves only to the terms that have the correct spatial wave
vector $\vec{k}_{\text{out}}=\vec{k}_1-\vec{k}_2+\vec{k}_{2^\prime}$. %
This approximation narrows the scope to sets of three interactions, one from each field, that
result in the correct spatial dependence. %
The set of three interactions have $3!=6$ unique time-ordered sequences, and each time-ordering
produces either two or three unique system-field interactions for our system, for a total of
sixteen unique system-field interaction sequences, or Liouville pathways, to consider. %
\autoref{mix:fig:WMELs} shows these sixteen pathways as Wave Mixing Energy Level (WMEL)
diagrams \cite{LeeDuckhwan1985a}. %
We first focus on a single interaction in these sequences, where an excitation pulse, $x$, forms
$\rho_{ij}$ from $\rho_{ik}$ or $\rho_{kj}$. %
For brevity, we use $\hbar=1$ and abbreviate the initial and final density matrix elements as
$\rho_i$ and $\rho_f$, respectively. %
Using the natural frequency rotating frame, $\tilde{\rho}_{ij}=\rho_{ij} e^{-i\omega_{ij}t}$, the formation of $\rho_f$ using pulse $x$ is written as
\begin{equation} \label{mix:eqn:rho_f}
\begin{split}
\frac{d\tilde{\rho}_f}{dt} =& -\Gamma_f\tilde{\rho}_f \\
&+ \frac{i}{2} \lambda_f \mu_f c_x(t-\tau_x)e^{i\kappa_f\left(\vec{k}_x\cdot z + \omega_x \tau_x \right)}e^{i\kappa_f\Omega_{fx}t}\tilde{\rho}_i(t),
\end{split}
\end{equation}
where $\Omega_{fx}=\kappa_f^{-1}\omega_f - \omega_x (=\left|\omega_f\right| - \omega_x)$ is the
detuning, $\omega_f$ is the transition frequency of the $i^{th}$ transition, $\mu_f$ is the
transition dipole, and $\Gamma_f$ is the dephasing/relaxation rate for $\rho_f$. %
The $\lambda_f$ and $\kappa_f$ parameters describe the phases of the interaction: $\lambda_f=+1$
for ket-side transitions and -1 for bra-side transitions, and $\kappa_f$ depends on whether
$\rho_f$ is formed via absorption ($\kappa_f= \lambda_f$) or emission
($\kappa_f=-\lambda_f$). %
$\kappa_f$ also has a direct relationship to the phase matching relationship: for transitions with
$E_2$, $\kappa_f=1$, and for $E_1$ or $E_{2^\prime}$, $\kappa_f=-1$. %
In the following equations we neglect spatial dependence ($z=0$). %
\autoref{mix:eqn:rho_f} forms the basis for our simulations. %
It provides a general expression for arbitrary values of the dephasing rate and excitation pulse
bandwidth. %
The integral solution is
\begin{equation} \label{mix:eqn:rho_f_int}
\begin{split}
\tilde{\rho}_f(t) =& \frac{i}{2}\lambda_f \mu_f e^{i\kappa_f \omega_x \tau_x} e^{i\kappa_f \Omega_{fx} t} \\
&\times \int_{-\infty}^{\infty} c_x(t-u-\tau_x)\tilde{\rho}_i(t-u)\Theta(u) \\
& \qquad \quad \ \ \times e^{-\left(\Gamma_f+i\kappa_f\Omega_{fx}\right)u}du,
\end{split}
\end{equation}
where $\Theta$ is the Heaviside step function. %
\autoref{mix:eqn:rho_f_int} becomes the steady state limit expression when $\Delta_t \left|\Gamma_f
+ i \kappa_f \Omega_{fx}\right| \gg 1$, and the impulsive limit expression results when $\Delta_t
\left|\Gamma_f + i \kappa_f \Omega_{fx}\right| \ll 1$. %
Both limits are important for understanding the multidimensional line shape changes discussed in
this paper. %
%The steady state and impulsive limits of Equation propagates \auotoref{mix:eqn:rho_f_int} are
% discussed in TODO
% Appendix \ref{sec:cw_imp}. %
\begin{figure}
\includegraphics[width=\linewidth]{"mixed_domain/simulation overview"}
\caption[Overview of the MR-CMDS simulation.]{
Overview of the MR-CMDS simulation.
(a) The temporal profile of a coherence under pulsed excitation depends on how quickly the
coherence dephases. In all subsequent panes, the relative dephasing rate is kept constant at
$\Gamma_{10}\Delta_t=1$.
(b) Simulated evolution of the density matrix elements of a third-order Liouville pathway
$V\gamma$ under fully resonant excitation. Pulses can be labeled both by their time of arrival
($x$,$y$,$z$) and by the lab lasers used to stimulate the transitions ($2$,$2^\prime$,$1$). The
final coherence (teal) creates the output electric field.
(c) The frequency profile of the output electric field is filtered by a monochromator gating
function, $M(\omega)$, and the passed components (shaded) are measured.
(d-f) Signal is viewed against two laser parameters, either as 2D delay (d), mixed
delay-frequency (e), or 2D frequency plots (f). The six time-orderings are labeled in (d) to
help introduce our delay convention.
}
\label{mix:fig:overview}
\end{figure}
\autoref{mix:fig:overview} gives an overview of the simulations done in this work. %
\autoref{mix:fig:overview}a shows an excitation pulse (gray-shaded) and examples of a coherent
transient for three different dephasing rates. %
The color bindings to dephasing rates introduced in \autoref{mix:fig:overview}a will be used
consistently throughout this work. %
Our simulations use systems with dephasing rates quantified relative to the pulse duration:
$\Gamma_{10} \Delta_t = 0.5, 1$, or $2$. %
The temporal axes are normalized to the pulse duration, $\Delta_t$. The $\Gamma_{10}\Delta_t=2$
transient is mostly driven by the excitation pulse while $\Gamma_{10} \Delta_t = 0.5$ has a
substantial free induction decay (FID) component at late times. %
\autoref{mix:fig:overview}b shows a pulse sequence (pulses are shaded orange and pink) and the
resulting system evolution of pathway $V\gamma$ ($00 \xrightarrow{2} 01 \xrightarrow{2^\prime} 11
\xrightarrow{1} 10 \xrightarrow{\text{out}} 00$) with $\Gamma_{10}\Delta_t=1$. %
The final polarization (teal) is responsible for the emitted signal, which is then passed through a
frequency bandpass filter to emulate monochromator detection (\autoref{mix:fig:overview}c). %
The resulting signal is explored in 2D delay space (\autoref{mix:fig:overview}d), 2D frequency space
(\autoref{mix:fig:overview}f), and hybrid delay-frequency space (\autoref{mix:fig:overview}e). %
The detuning frequency axes are also normalized by the pulse bandwidth, $\Delta_{\omega}$. %
We now consider the generalized Liouville pathway $L:\rho_0 \xrightarrow{x} \rho_1 \xrightarrow{y}
\rho_2 \xrightarrow{z} \rho_3 \xrightarrow{\text{out}} \rho_4$, where $x$, $y$, and $z$ denote
properties of the first, second, and third pulse, respectively, and indices 0, 1, 2, 3, and 4
define the properties of the ground state, first, second, third, and fourth density matrix
elements, respectively. %
\autoref{mix:fig:overview}b demonstrates the correspondence between $x$, $y$, $z$ notation and 1, 2,
$2^\prime$ notation for the laser pulses with pathway $V\gamma$.
The electric field emitted from a Liouville pathway is proportional to the polarization created by
the third-order coherence: %
\begin{equation} \label{mix:eqn:E_L}
E_L(t) = i \mu_{4}\rho_{3}(t).
\end{equation}
\autoref{mix:eqn:E_L} assumes perfect phase-matching and no pulse distortions through propagation.
\autoref{mix:eqn:rho_f_int} shows that the output field for this Liouville pathway is
\begin{gather} \label{mix:eqn:E_L_full}
\begin{split}
E_L(t) =& \frac{i}{8}\lambda_1\lambda_2\lambda_3\mu_1\mu_2\mu_3\mu_4
e^{i\left( \kappa_1\omega_x\tau_x + \kappa_2\omega_y\tau_y + \kappa_3\omega_z\tau_z \right)}
e^{-i\left( \kappa_3 \omega_z + \kappa_2 \omega_y + \kappa_1 \omega_x \right) t} \\
&\times \iiint_{-\infty}^{\infty} c_z(t-u-\tau_z) c_y(t-u-v-\tau_y) c_x(t-u-v-w-\tau_x) R_L(u,v,w) dw \ dv \ du ,
\end{split}\\
R_L(u,v,w) = \Theta(w)e^{-\left(\Gamma_1 + i\kappa_1\Omega_{1x} \right)w}
\Theta(v)e^{-\left(\Gamma_2 + i\left[ \kappa_1\Omega_{1x}+\kappa_2\Omega_{2y} \right] \right)v}
\Theta(u)e^{-\left(\Gamma_3 + i\left[ \kappa_1\Omega_{1x}+\kappa_2\Omega_{2y}+\kappa_3\Omega_{3z} \right] \right)u},
\end{gather}
where $R_L$ is the third-order response function for the Liouville pathway. %
The total electric field will be the superposition of all the Liouville pathways:
\begin{equation} \label{mix:eqn:superposition}
E_{\text{tot}}= \sum_L E_L(t).
\end{equation}
For the superposition of \autoref{mix:eqn:superposition} to be non-canceling, certain symmetries
between the pathways must be broken. %
In general, this requires one or more of the following inequalities: $\Gamma_{10}\neq\Gamma_{21}$,
$\omega_{10}\neq\omega_{21}$, and/or $\sqrt{2}\mu_{10}\neq\mu_{21}$. %
Our simulations use the last inequality, which is important in two-level systems ($\mu_{21}=0$) and
in systems where state-filling dominates the non-linear response, such as in semiconductor
excitons. %
The exact ratio between $\mu_{10}$ and $\mu_{21}$ affects the absolute amplitude of the field, but
does not affect the multidimensional line shape. %
Importantly, the dipole inequality does not break the symmetry of double quantum coherence pathways
(time-orderings II and IV), so such pathways are not present in our analysis. %
In MR-CMDS, a monochromator resolves the driven output frequency from other nonlinear output
frequencies, which in our case is $\omega_m = \omega_1 - \omega_2 + \omega_{2'} = \omega_1$. %
The monochromator can also enhance spectral resolution, as we show in
\autoref{mix:sec:evolution_SQC}. %
In this simulation, the detection is emulated by transforming $E_{\text{tot}}(t)$ into the
frequency domain, applying a narrow bandpass filter, $M(\omega)$, about $\omega_1$, and applying
amplitude-scaled detection:
\begin{equation} \label{mix:eqn:S_tot}
S_{\text{tot}}(\omega_1, \omega_2, \tau_{21}, \tau_{22^\prime}) =
\sqrt{ \int\left| M(\omega-\omega_1) E_{\text{tot}}(\omega) \right|^2 d\omega},
\end{equation}
where $E_{\text{tot}}(\omega)$ denotes the Fourier transform of $E_{\text{tot}}(t)$ (see
\autoref{mix:fig:overview}c). %
For $M$ we used a rectangular function of width $0.408\Delta_{\omega}$. %
The arguments of $S_{\text{tot}}$ refer to the \textit{experimental} degrees of freedom. %
The signal delay dependence is parameterized with the relative delays $\tau_{21}$ and
$\tau_{22^\prime}$, where $\tau_{nm} = \tau_n - \tau_m$ (see \autoref{mix:fig:overview}b). %
Table S1 summarizes the arguments for each Liouville pathway. %
\autoref{mix:fig:overview}f shows the 2D $(\omega_1, \omega_2)$ $S_{\text{tot}}$ spectrum resulting
from the pulse delay times represented in \autoref{mix:fig:overview}b. %
\subsection{Inhomogeneity} % ---------------------------------------------------------------------
Inhomogeneity is isolated in CMDS through both spectral signatures, such as
line-narrowing \cite{BesemannDanielM2004a, OudarJL1980a, CarlsonRogerJohn1990a, RiebeMichaelT1988a,
SteehlerJK1985a}, and temporal signatures, such as photon echoes \cite{WeinerAM1985a,
AgarwalRitesh2002a}. %
We simulate the effects of static inhomogeneous broadening by convolving the homogeneous response
with a Gaussian distribution function. %
Further details of the convolution are in \autoref{mix:sec:convolution}. %
Dynamic broadening effects such as spectral diffusion are beyond the scope of this work. %
\section{Methods} % ==============================================================================
A matrix representation of differential equations of the type in \autoref{mix:eqn:E_L_full} was
numerically integrated for parallel computation of Liouville elements (see SI for
details). \cite{DickBernhard1983a, GelinMaximF2005a} %
The lower bound of integration was $2\Delta_t$ before the first pulse, and the upper bound was
$5\Gamma_{10}^{-1}$ after the last pulse, with step sizes much shorter than the pulse durations. %
Integration was performed in the FID rotating frame; the time steps were chosen so that both the
system-pulse difference frequencies and the pulse envelope were well-sampled. %
The following simulations explore the four-dimensional $(\omega_1, \omega_2, \tau_{21},
\tau_{22^\prime})$ variable space. %
Both frequencies are scanned about the resonance, and both delays are scanned about pulse overlap.
We explored the role of sample dephasing rate by calculating signal for systems with dephasing
rates such that $\Gamma_{10}\Delta_t=0.5, 1,$ and $2$. %
Inhomogeneous broadening used a spectral FWHM, $\Delta_{\text{inhom}}$, that satisfied
$\Delta_\text{inhom}/ \Delta_{\omega}=0,0.5,1,$ and $2$ for the three dephasing rates. %
For all these dimensions, both $\rho_3(t;\omega_1, \omega_2, \tau_{21}, \tau_{22^\prime})$ and
$S_{\text{tot}}(\omega_1, \omega_2, \tau_{21}, \tau_{22^\prime})$ are recorded for each unique
Liouville pathway. %
Our simulations were done using the open-source SciPy library. \cite{OliphantTravisE2007a} %
\subsection{Characteristics of Driven and Impulsive Response} \label{mix:sec:cw_imp} % -----------
The changes in the spectral line shapes described in this work are best understood by examining the
driven/continuous wave (CW) and impulsive limits of \autoref{mix:eqn:rho_f_int} and
\ref{mix:eqn:E_L_full}. %
The driven limit is achieved when pulse durations are much longer than the response function
dynamics: $\Delta_t \left|\Gamma_f + i \kappa_f \Omega_{fx}\right| \gg 1$. %
In this limit, the system will adopt a steady state over excitation: $d\rho / dt \approx 0$. %
Neglecting phase factors, the driven solution to \autoref{mix:eqn:rho_f_int} will be
\begin{equation} \label{mix:eqn:sqc_driven}
\tilde{\rho}_f(t) = \frac{\lambda_f \mu_f}{2}
\frac{c_x(t-\tau_x)e^{i\kappa_f \Omega_{fx}t}}{\kappa_f \Omega_{fx}} \tilde{\rho}_i(t).
\end{equation}
The frequency and temporal envelope of the excitation pulse controls the coherence time evolution,
and the relative amplitude and phase of the coherence is directly related to detuning from
resonance. %
The impulsive limit is achieved when the excitation pulses are much shorter than response function
dynamics: $\Delta_t \left|\Gamma_f + i \kappa_f \Omega_{fx}\right| \ll 1$. %
The full description of the temporal evolution has two separate expressions: one for times when the
pulse is interacting with the system, and one for times after pulse interaction. %
Both expressions are important when describing CMDS experiments. %
For times after the pulse interaction, $t \gtrsim \tau_x + \Delta_t$, the field-matter coupling is
negligible. %
The evolution for these times, on resonance, is given by
\begin{equation} \label{mix:eqn:sqc_fid}
\tilde{\rho}_f(t) =\frac{i \lambda_f\mu_f }{2} \tilde{\rho}_i(\tau_x)
\int c_x(u) du \ e^{-\Gamma_f(t-\tau_x)}.
\end{equation}
This is classic free induction decay (FID) evolution: the system evolves at its natural frequency
and decays at rate $\Gamma_f$. %
It is important to note that, while this expression is explicitly derived from the impulsive limit,
FID behavior is not exclusive to impulsive excitation, as we have defined it. %
A latent FID will form if the pulse vanishes at a fast rate relative to the system dynamics.
For evaluating times near pulse excitation, $t \lesssim \tau_x + \Delta_t$, we implement a Taylor
expansion in the response function about zero: $e^{-(\Gamma_f+i\kappa_f\Omega_{fx})u} = 1 -
(\Gamma_f+i\kappa_f\Omega_{fx})u+\cdots$. %
Our impulsive criterion requires that a low order expansion will suffice; it is instructive to
consider the result of the first order expansion of \autoref{mix:eqn:rho_f_int}: %
\begin{equation} \label{mix:eqn:sqc_rise}
\begin{split}
\tilde{\rho}_f(t) =& \frac{i \lambda_f\mu_f}{2} e^{-i\kappa_f\omega_x\tau_x}e^{-i\kappa_f\Omega_{fx}t} \tilde{\rho}_i(\tau_x) \\
& \times \bigg[ \left( 1-(\Gamma_f + i\kappa_f\Omega_{fx})(t-\tau_x) \right) \int_{-\infty}^{t-\tau_x} c_x(u) du \\
& \quad +(\Gamma_f + i\kappa_f\Omega_{fx}) \int_{-\infty}^{t-\tau_x} c_x(u)u \ du \bigg].
\end{split}
\end{equation}
During this time $\tilde{\rho}_f$ builds up roughly according to the integration of the pulse
envelope. %
The build-up is integrated because the pulse transfers energy before appreciable dephasing or
detuning occurs. %
Contrary to the expectation of impulsive evolution, the evolution of $\tilde{\rho}_f$ is explicitly
affected by the pulse frequency, and the temporal profile evolves according to the pulse. %
It is important to recognize that the impulsive limit is defined not only by having slow relaxation
relative to the pulse duration, but also by small detuning relative to the pulse bandwidth (as is
stated in the inequality). %
As detuning increases, the higher orders of the response function Taylor expansion will be needed
to describe the rise time, and the driven limit of \autoref{mix:eqn:sqc_driven} will become
valid. %
The details of this build-up time can often be neglected in impulsive approximations because
build-up contributions are often negligible in analysis; the period over which the initial
excitation occurs is small in comparison to the free evolution of the system. %
The build-up behavior can be emphasized by the measurement, which makes \autoref{mix:eqn:sqc_rise}
important. %
We now consider full Liouville pathways in the impulsive and driven limits of
\autoref{mix:eqn:E_L_full}. %
For the driven limit, \autoref{mix:eqn:E_L_full} can be reduced to
\begin{equation} \label{mix:eqn:E_L_driven}
\begin{split}
E_L(t) =& \frac{1}{8} \lambda_1\lambda_2\lambda_3\mu_1\mu_2\mu_3\mu_4
e^{-i(\kappa_1\omega_x\tau_x + \kappa_2\omega_y\tau_y + \kappa_3\omega_z\tau_z)} \\
& \times e^{ i(\kappa_3\omega_z + \kappa_2\omega_y + \kappa_1\omega_x)t} \\
& \times c_z(t-\tau_z)c_y(t-\tau_y)c_x(t-\tau_x) \\
& \times \frac{1}{\kappa_1\Omega_{1x}-i\Gamma_1} \frac{1}{\kappa_1\Omega_{1x} + \kappa_2\Omega_{2y} - i\Gamma_2} \\
& \times \frac{1}{\kappa_1\Omega_{1x} + \kappa_2 \Omega_{2y} + \kappa_3\Omega_{3z}-i\Gamma_3}.
\end{split}
\end{equation}
It is important to note that the signal depends on the multiplication of all the fields; pathway
discrimination based on pulse time-ordering is not achievable because polarizations exists only
when all pulses are overlapped. %
This limit is the basis for frequency-domain techniques. %
Frequency axes, however, are not independent because the system is forced to the laser frequency
and influences the resonance criterion for subsequent excitations. %
As an example, observe that the first two resonant terms in \autoref{mix:eqn:E_L_driven} are
maximized when $\omega_x=\left|\omega_1\right|$ and $\omega_y=\left|\omega_2\right|$. %
If $\omega_x$ is detuned by some value $\varepsilon$, however, the occurrence of the second
resonance shifts to $\omega_y=\left|\omega_2\right|+\varepsilon$, effectively compensating for the
$\omega_x$ detuning. %
This shifting of the resonance results in 2D line shape correlations. %
If the pulses do not temporally overlap $(\tau_x+\Delta_t \lesssim \tau_y +\Delta_t \lesssim \tau_z
+ \Delta_t \lesssim t)$, then the impulsive solution to the full Liouville pathway of
\autoref{mix:eqn:E_L_full} is %
\begin{equation} \label{mix:eqn:E_L_impulsive}
\begin{split}
E_L(t) =& \frac{i}{8} \lambda_1\lambda_2\lambda_3\mu_1 \mu_2 \mu_3 \mu_4 e^{i(\omega_1 + \omega_2 + \omega_3)t} \\
& \times \int c_x(w) dw \int c_y(v) dv \int c_z(u) du \\
& \times e^{-\Gamma_1(\tau_y-\tau_x)} e^{-\Gamma_2(\tau_z-\tau_y)} e^{-\Gamma(t-\tau_z)}.
\end{split}
\end{equation}
Pathway discrimination is demonstrated here because the signal is sensitive to the time-ordering of
the pulses. %
This limit is suited for delay scanning techniques. %
The emitted signal frequency is determined by the system and can be resolved by scanning a
monochromator. %
The driven and impulsive limits can qualitatively describe our simulated signals at certain
frequency and delay combinations. %
Of the three expressions, the FID limit most resembles signal when pulses are near resonance and
well-separated in time (so that build-up behavior is negligible). %
The build-up limit approximates well when pulses are near-resonant and arrive together (so that
build-up behavior is emphasized). %
The driven limit holds for large detunings, regardless of delay. %
\subsection{Convolution technique for inhomogeneous broadening} \label{mix:sec:convolution} % ----
\begin{figure}
\includegraphics[width=\linewidth]{mixed_domain/convolve}
\caption[Convolution overview.]
{Overview of the convolution.
(a) The homogeneous line shape.
(b) The distribution function, $K$, mapped onto laser coordinates.
(c) The resulting ensemble line shape computed from the convolution.
The thick black line represents the FWHM of the distribution function.}
\label{mix:fig:convolution}
\end{figure}
Here we describe how to transform the data of a single reference oscillator signal to that of an
inhomogeneous distribution. %
The oscillators in the distribution are allowed have arbitrary energies for their states, which
will cause frequency shifts in the resonances. %
To show this, we start with a modified, but equivalent, form of \autoref{mix:eqn:rho_f}:
\begin{equation} \label{mix:eqn:rho_f_modified}
\begin{split}
\frac{d\tilde{\rho}_f}{dt} =& -\Gamma_f\tilde{\rho}_f + \frac{i}{2}\lambda_f\mu_f c_x(t-\tau_x) \\
& \times e^{i\kappa_f\left( \vec{k}\cdot z + \omega_x \tau_x \right)} e^{-i\kappa_f\left( \omega_x-\left|\omega_f \right| \right)t}\tilde{\rho}_i(t).
\end{split}
\end{equation}
We consider two oscillators with transition frequencies $\omega_f$ and $\omega_f^\prime=\omega_f +
\delta$. %
So long as $\left| \delta \right| \leq \omega_f$ (so that $\left| \omega_f + \delta \right| =
\left| \omega_f \right| + \delta$ and thus the rotating wave approximation does not change),
\autoref{mix:eqn:rho_f_modified} shows that the two are related by %
\begin{equation} \label{mix:eqn:freq_translation}
\frac{d\tilde{\rho}_f^\prime}{dt}(t;\omega_x) = \frac{d\tilde{\rho}_f}{dt}(t;\omega_x-\delta)e^{i\kappa_f \delta \tau_x}.
\end{equation}
Because both coherences are assumed to have the same initial conditions
($\rho_0(-\infty)=\rho_0^\prime(-\infty)=0$), the equality also holds when both sides of the
equation are integrated. %
The phase factor $e^{i\kappa_f\delta\tau_x}$ in the substitution arises from \autoref{mix:eqn:E_l},
where the pulse carrier frequency maintains its phase within the pulse envelope for all delays. %
The resonance translation can be extended to higher order signals as well. %
For a third-order signal, we compare systems with transition frequencies
$\omega_{10}^\prime=\omega_{10}+a$ and $\omega_{21}^\prime = \omega_{21}+b$. %
The extension of \autoref{mix:eqn:freq_translation} to pathway $V\beta$ gives %
\begin{equation}
\begin{split}
\tilde{\rho}_3^\prime(t;\omega_2, \omega_2^\prime, \omega_1) =& \tilde{\rho}_3(t;\omega_2-a,\omega_{2^\prime}-a,\omega_1-b) \\
&\times e^{i\kappa_2 a \tau_2} e^{i\kappa_{2^\prime} a \tau_{2^\prime}} e^{i\kappa_1 b \tau_1}.
\end{split}
\end{equation}
The translation of each laser coordinate depends on which transition is made (e.g. $a$ for
transitions between $|0\rangle$ and $|1\rangle$ or $b$ for transitions between $|1\rangle$ and
$|2\rangle$), so the exact translation relation differs between pathways. %
We can now compute the ensemble average of signal for pathway $V\beta$ as a convolution between the
distribution function of the system, $K(a,b)$, and the single oscillator response: %
\begin{equation}
\begin{split}
\langle \tilde{\rho}_3 (t;\omega_2,\omega_{2^\prime},\omega_1) \rangle =& \iint K(a,b)\\
& \times \tilde{\rho}_3 (t;\omega_2+a,\omega_{2^\prime}+a,\omega_1+b) \\
& \times e^{i\kappa_2 a \tau_2} e^{i\kappa_{2^\prime} a \tau_{2^\prime}} e^{i\kappa_1 b \tau_1} da \ db.
\end{split}
\end{equation}
For this work, we restrict ourselves to a simpler ensemble where all oscillators have equally
spaced levels (i.e. $a=b$). %
This makes the translation identical for all pathways and reduces the dimensionality of the
convolution. %
Since pathways follow the same convolution we may also perform the convolution on the total signal field:
\begin{equation}
\begin{split}
\langle E_{\text{tot}}(t) \rangle =& \sum_L \mu_{4,L} \int K(a,a) \\
& \times \tilde{\rho}_{3,L}(t;\omega_x-a,\omega_y-a\omega_z-a) \\
& \times e^{ia\left( \kappa_x\tau_x + \kappa_y\tau_y + \kappa_z\tau_z \right)} da.
\end{split}
\end{equation}
Furthermore, since $\kappa=-1$ for $E_1$ and $E_{2^\prime}$, while $\kappa=1$ for $E_2$, we have
$e^{ia\left( \kappa_x\tau_x + \kappa_y\tau_y + \kappa_z\tau_z \right)} = e^{-ia\left( \tau_1 -
\tau_2 + \tau_{2^\prime} \right)}$ for all pathways. %
Equivalently, if the electric field is parameterized in terms of laser coordinates $\omega_1$ and $\omega_2$, the ensemble field can be calculated as
\begin{equation} \label{mix:eqn:convolve_final}
\begin{split}
\langle E_{\text{tot}}(t;\omega_1,\omega_2) \rangle =& \int K(a,a)E_{\text{tot}}(t;\omega_1-a,\omega_2-a) \\
&\times e^{-ia\left( \tau_1-\tau_2+\tau_{2^\prime} \right)} da.
\end{split}
\end{equation}
which is a 1D convolution along the diagonal axis in frequency space. %
\autoref{mix:fig:convolution} demonstrates the use of \autoref{mix:eqn:convolve_final} on a
homogeneous line shape. %
\section{Results} % ==============================================================================
We now present portions of our simulated data that highlight the dependence of the spectral line
shapes and transients on excitation pulse width, the dephasing rate, detuning from resonance, the
pulse delay times, and inhomogeneous broadening. %
\subsection{Evolution of single coherence} \label{mix:sec:evolution_SQC} % -----------------------
\begin{figure}
\centering
\includegraphics[width=0.5\linewidth]{"mixed_domain/fid vs dpr"}
\caption[Relative importance of FID and driven response for a single quantum coherence.]{
The relative importance of FID and driven response for a single quantum coherence as a function
of the relative dephasing rate (values of $\Gamma_{10}\Delta_t$ are shown inset).
The black line shows the coherence amplitude profile, while the shaded color indicates the
instantaneous frequency (see colorbar).
For all cases, the pulsed excitation field (gray line, shown as electric field amplitude) is
slightly detuned (relative detuning, $\Omega_{fx}/\Delta_{\omega}=0.1$).
}
\label{mix:fig:fid_dpr}
\end{figure}
It is illustrative to first consider the evolution of single coherences, $\rho_0 \xrightarrow{x}
\rho_1$, under various excitation conditions. %
\autoref{mix:fig:fid_dpr} shows the temporal evolution of $\rho_1$ with various dephasing rates under
Gaussian excitation. %
The value of $\rho_1$ differs only by phase factors between various Liouville pathways (this can be
verified by inspection of \ref{mix:eqn:rho_f_int} under the various conditions in Table S1), so
the profiles in \autoref{mix:fig:fid_dpr} apply for the first interaction of any pathway. %
The pulse frequency was detuned from resonance so that frequency changes could be visualized by the
color bar, but the detuning was kept slight so that it did not appreciably change the dimensionless
product, $\Delta_t \left(\Gamma_f + i\kappa_f \Omega_{fx}\right)\approx \Gamma_{10}\Delta_t$. %
In this case, the evolution demonstrates the maximum impulsive character the transient can
achieve. %
The instantaneous frequency, $d\varphi/dt$, is defined as
\begin{equation}
\frac{d\varphi}{dt} = \frac{d}{dt} \tan^{-1}\left( \frac{\text{Im}\left(\rho_1(t)\right)}{\text{Re}\left(\rho_1(t)\right)} \right).
\end{equation}
The cases of $\Gamma_{10}\Delta_t=0 (\infty)$ agree with the impulsive (driven) expressions derived
in \autoref{mix:sec:cw_imp}. %
For $\Gamma_{10}\Delta_t=0$, the signal rises as the integral of the pulse and has instantaneous
frequency close to that of the pulse (\autoref{mix:eqn:sqc_rise}), but as the pulse vanishes, the
signal adopts the natural system frequency and decay rate (\autoref{mix:eqn:sqc_fid}). %
For $\Gamma_{10}\Delta_t=\infty$, the signal follows the amplitude and frequency of the pulse for
all times (the driven limit, \autoref{mix:eqn:sqc_driven}). %
The other three cases show a smooth interpolation between limits. %
As $\Gamma_{10}\Delta_t$ increases from the impulsive limit, the coherence within the pulse region
conforms less to a pulse integral profile and more to a pulse envelope profile. %
In accordance, the FID component after the pulse becomes less prominent, and the instantaneous
frequency pins to the driving frequency more strongly through the course of evolution. %
The trends can be understood by considering the differential form of evolution (
\autoref{mix:eqn:rho_f}), and the time-dependent balance of optical coupling and system relaxation. %
We note that our choices of $\Gamma_{10}\Delta_t=2.0, 1.0,$ and $0.5$ give coherences that have
mainly driven, roughly equal driven and FID parts, and mainly FID components, respectively. %
FID character is difficult to isolate when $\Gamma_{10}\Delta_t=2.0$. %
\begin{figure}
\centering
\includegraphics[width=0.5\linewidth]{"mixed_domain/fid vs detuning"}
\caption[Pulsed excitation of a single quantum coherence and its dependance on pulse detuning.]{
Pulsed excitation of a single quantum coherence and its dependence on the pulse detuning. In
all cases, relative dephasing is kept at $\Gamma_{10}\Delta_t = 1$.
(a) The relative importance of FID and driven response for a single quantum coherence as a
function of the detuning (values of relative detuning, $\Omega_{fx}/\Delta_{\omega}$, are shown
inset).
The color indicates the instantaneous frequency (scale bar on right), while the black line
shows the amplitude profile. The gray line is the electric field amplitude.
%Comparison of the temporal evolution of single quantum coherences at different detunings
%(labeled inset).
(b) The time-integrated coherence amplitude as a function of the detuning. The integrated
amplitude is collected both with (teal) and without (magenta) a tracking monochromator that
isolates the driven frequency components. %$\omega_m=\omega_\text{laser}$.
For comparison, the Green's function of the single quantum coherence is also shown (amplitude
is black, hashed; imaginary is black, solid).
In all plots, the gray line is the electric field amplitude.
}
\label{mix:fig:fid_detuning}
\end{figure}
\autoref{mix:fig:fid_detuning}a shows the temporal evolution of $\rho_1$ at several values of
$\Omega_{1x}/\Delta_{\omega}$ with $\Gamma_{10}\Delta_t=1$.
As detuning increases, total amplitude decreases, FID character vanishes, and $\rho_1$ assumes a
more driven character, as expected. %
During the excitation, $\rho_1$ maintains a phase relationship with the input field (as seen by the
instantaneous frequency in \autoref{mix:fig:fid_detuning}a). %
The coherence will persist beyond the pulse duration only if the pulse transfers energy into the
system; FID evolution equates to absorption. %
The FID is therefore sensitive to the absorptive (imaginary) line shape of a transition, while the
driven response is the composite of both absorptive and dispersive components. %
If the experiment isolates the latent FID response, there is consequently a narrower spectral
response. %
This spectral narrowing can be seen in \autoref{mix:fig:fid_detuning}a by comparing the coherence
amplitudes at $t=0$ (driven) and at $t / \Delta_t=2$ (FID); amplitudes for all
$\Omega_{fx}/\Delta_{\omega}$ values shown are comparable at $t=0$, but the lack of FID formation
for $\Omega_{fx}/\Delta_{\omega}=\pm2$ manifests as a visibly disproportionate amplitude decay. %
Many ultrafast spectroscopies take advantage of the latent FID to suppress non-resonant background,
improving signal to noise. \cite{LagutchevAlexi2007a, LagutchevAlexi2010a,
DonaldsonPaulMurray2010a, DonaldsonPaulMurray2008a} %
In driven experiments, the output frequency and line shape are fully constrained by the excitation
beams. %
In such experiments, there is no additional information to be resolved in the output spectrum. %
The situation changes in the mixed domain, where $E_\text{tot}$ contains FID signal that lasts
longer than the pulse duration. %
\autoref{mix:fig:fid_detuning}a provides insight on how frequency-resolved detection of coherent
output can enhance resolution when pulses are spectrally broad. %
Without frequency-resolved detection, mixed-domain resonance enhancement occurs in two ways: (1)
the peak amplitude increases, and (2) the coherence duration increases due to the FID transient. %
Frequency-resolved detection can further discriminate against detuning by requiring that the
driving frequency agrees with latent FID. %
The implications of discrimination are most easily seen in \autoref{mix:fig:fid_detuning}a with
$\Omega_{1x}/\Delta_{\omega}=\pm 1$, where the system frequency moves from the driving frequency to
the FID frequency. %
When the excitation pulse frequency is scanned, the resonance will be more sensitive to detuning by
isolating the driven frequency (tracking the monochromator with the excitation source). %
The functional form of the measured line shape can be deduced by considering the frequency domain
form of \autoref{mix:eqn:rho_f_int} (assume $\rho_i=1$ and $\tau_x=0$):
\begin{equation} \label{mix:eqn:rho_f_int_freq}
\tilde{\rho}_f (\omega) = \frac{i\lambda_f\mu_f}{2\sqrt{2\pi}} \cdot \frac{\mathcal{F}\left\{ c_x \right\}\left( \omega - \kappa_f\Omega_{fx} \right)}{\Gamma_f + i\omega},
\end{equation}
where $\mathcal{F}\left\{ c_x \right\}\left( \omega \right)$ denotes the Fourier transform of
$c_x$, which in our case gives
\begin{equation}
\mathcal{F}\left\{ c_x \right\}\left( \omega \right) = \frac{1}{\sqrt{2\pi}} e^{-\frac{(\Delta_t\omega)^2}{4\ln 2}}.
\end{equation}
For squared-law detection of $\rho_f$, the importance of the tracking monochromator is highlighted
by two limits of \autoref{mix:eqn:rho_f_int_freq}:
\begin{itemize}
\item When the transient is not frequency resolved, $\text{sig} \approx \int{\left|
\tilde{\rho}_f(\omega) \right|^2 d\omega}$ and the measured line shape will be the
convolution of the pulse envelope and the intrinsic (Green's function) response
(\autoref{mix:fig:fid_detuning}b, magenta).
\item When the driven frequency is isolated, $\text{sig} \approx \left|
\tilde{\rho}_f(\kappa_f\Omega_{fx}) \right|^2$ and the measured line shape will give the
un-broadened Green's function (\autoref{mix:fig:fid_detuning}b, teal).
\end{itemize}
Monochromatic detection can remove broadening effects due to the pulse bandwidth. %
For large $\Gamma_{10}\Delta_t$ values, FID evolution is negligible at all
$\Omega_{fx}/\Delta_{\omega}$ values and the monochromator is not useful. %
\autoref{mix:fig:fid_detuning}b shows the various detection methods for the relative dephasing rate of
$\Gamma_{10}\Delta_t=1$. %
\subsection{Evolution of single Liouville pathway} % ---------------------------------------------
\begin{figure}
\centering
\includegraphics[width=\linewidth]{"mixed_domain/pw1 lineshapes"}
\caption[2D frequency response of a single Liouville pathway at different delay values.]{
Changes to the 2D frequency response of a single Liouville pathway (I$\gamma$) at different
delay values.
The normalized dephasing rate is $\Gamma_{10}\Delta_t = 1$.
Left: The 2D delay response of pathway I$\gamma$ at triple resonance.
Right: The 2D frequency response of pathway I$\gamma$ at different delay values.
The delays at which the 2D frequency plots are collected are indicated on the delay plot;
compare 2D spectrum frame color with dot color on 2D delay plot.
}
\label{mix:fig:pw1}
\end{figure}
We now consider the multidimensional response of a single Liouville pathway involving three pulse
interactions. %
In a multi-pulse experiment, $\rho_1$ acts as a source term for $\rho_2$ (and subsequent
excitations). %
The spectral and temporal features of $\rho_1$ that are transferred to $\rho_2$ depend on when the
subsequent pulse arrives. %
Time-gating later in $\rho_1$ evolution will produce responses with FID behavior, while time-gating
$\rho_1$ in the presence of the initial pulse will produce driven responses. %
An analogous relationship holds for $\rho_3$ with its source term $\rho_2$. %
As discussed above, signal that time-gates FID evolution gives narrower spectra than driven-gated
signal. %
As a result, the spectra of even single Liouville pathways will change based on pulse delays. %
The final coherence will also be frequency-gated by the monochromator. %
The monochromator isolates signal at the fully driven frequency $\omega_\text{out} = \omega_1$. %
The monochromator will induce line-narrowing to the extent that FID takes place. %
It effectively enforces a frequency constraint that acts as an additional resonance condition,
$\omega_\text{out}=\omega_1$. %
The driven frequency will be $\omega_1$ if $E_1$ is the last pulse interaction (time-orderings V
and VI), and the monochromator tracks the coherence frequency effectively. %
If $E_1$ is not the last interaction, the output frequency may not be equal to the driven
frequency, and the monochromator plays a more complex role. %
We demonstrate this delay dependence using the multidimensional response of the I$\gamma$ Liouville
pathway as an example (see \autoref{mix:fig:WMELs}). %
\autoref{mix:fig:pw1} shows the resulting 2D delay profile of pathway I$\gamma$ signals for
$\Gamma_{10}\Delta_t=1$ (left) and the corresponding $\omega_1, \omega_2$ 2D spectra at several
pulse delay values (right). %
The spectral changes result from changes in the relative importance of driven and FID
components. %
The prominence of FID signal can change the resonance conditions; \autoref{mix:tab:table2} summarizes
the changing resonance conditions for each of the four delay coordinates studied. %
Since $E_1$ is not the last pulse in pathway I$\gamma$, the tracking monochromator must also be
considered. %
\begin{table}
\caption{\label{mix:tab:table2} Conditions for peak intensity at different pulse delays for pathway
I$\gamma$.}
\begin{tabular}{c c | c c c c}
\multicolumn{2}{c}{Delay} & \multicolumn{4}{|c}{Approximate Resonance Conditions} \\
$\tau_{21}/\Delta_t$ & $\tau_{22^\prime}/\Delta_t$ & $\rho_0\xrightarrow{1}\rho_1$ &
$\rho_1\xrightarrow{2}\rho_2$ & $\rho_2\xrightarrow{2^\prime}\rho_3$ & $\rho_3\rightarrow$
detection at $\omega_m=\omega_1$ \\
\hline\hline
0 & 0 & $\omega_1=\omega_{10}$ & $\omega_1=\omega_2$ & $\omega_1=\omega_{10}$ & -- \\
0 & -2.4 & $\omega_1=\omega_{10}$ & $\omega_1=\omega_2$ & $\omega_2=\omega_{10}$ &
$\omega_1=\omega_2$ \\
2.4 & 0 & $\omega_1=\omega_{10}$ & $\omega_2=\omega_{10}$ & -- &
$\omega_1=\omega_{10}$ \\
2.4 & -2.4 & $\omega_1=\omega_{10}$ & $\omega_2=\omega_{10}$ & $\omega_2=\omega_{10}$ &
$\omega_1=\omega_2$ \\
\end{tabular}
\end{table}
When the pulses are all overlapped ($\tau_{21}=\tau_{22^\prime}=0$, lower right, orange), all
transitions in the Liouville pathway are simultaneously driven by the incident fields. %
This spectrum strongly resembles the driven limit spectrum. For this time-ordering, the first,
second, and third density matrix elements have driven resonance conditions of
$\omega_1=\omega_{10}$, $\omega_1-\omega_2=0$, and
$\omega_1-\omega_2+\omega_{2^\prime}=\omega_{10}$, respectively. %
The second resonance condition causes elongation along the diagonal, and since
$\omega_2=\omega_{2^\prime}$, the first and third resonance conditions are identical, effectively
making $\omega_1$ doubly resonant at $\omega_{10}$ and resulting in the vertical elongation along
$\omega_1=\omega_{10}$. %
The other three spectra of \autoref{mix:fig:pw1} separate the pulse sequence over time so that not all
interactions are driven. %
At $\tau_{21}=0$, $\tau_{22^\prime}=-2.4\Delta_t$ (lower left, pink), the first two resonances
remain the same as at pulse overlap (orange) but the last resonance is different. %
The final pulse, $E_{2^\prime}$, is latent and probes $\rho_2$ during its FID evolution after
memory of the driven frequency is lost. %
There are two important consequences. %
Firstly, the third driven resonance condition is now approximated by
$\omega_{2^\prime}=\omega_{10}$, which makes $\omega_1$ only singly resonant at
$\omega_1=\omega_{10}$. %
Secondly, the driven portion of the signal frequency is determined only by the latent pulse:
$\omega_{\text{out}}=\omega_{2^\prime}$. %
Since our monochromator gates $\omega_1$, we have the detection-induced correlation
$\omega_1=\omega_{2^\prime}$. %
The net result is double resonance along $\omega_1=\omega_2$, and the vertical elongation of pulse
overlap is strongly attenuated. %
At $\tau_{21}=2.4\Delta_t,\tau_{22^\prime}=0$ (upper right, purple), the first pulse $E_1$ precedes
the latter two, which makes the two resonance conditions for the input fields
$\omega_1=\omega_{10}$ and $\omega_2=\omega_{10}$. %
The signal depends on the FID conversion of $\rho_1$, which gives vertical elongation at
$\omega_1=\omega_{10}$. %
Furthermore, $\rho_1$ has no memory of $\omega_1$ when $E_2$ interacts, which has two important
implications. %
First, this means the second resonance condition $\omega_1=\omega_2$ and the associated diagonal
elongation is now absent. %
Second, the final output polarization frequency content is no longer functional of $\omega_1$.
Coupled with the fact that $E_2$ and $E_{2^\prime}$ are coincident, so that the final coherence can
be approximated as driven by these two, we can approximate the final frequency as
$\omega_{\text{out}} = \omega_{10}-\omega_2+\omega_{2^\prime} = \omega_{10}$. %
Surprisingly, the frequency content of the output is strongly independent of all pulse
frequencies. %
The monochromator narrows the $\omega_1=\omega_{10}$ resonance. %
The $\omega_1=\omega_{10}$ resonance condition now depends on the monochromator slit width, the FID
propagation of $\rho_1$, the spectral bandwidth of $\rho_3$; its spectral width is not easily
related to material parameters. %
This resonance demonstrates the importance of the detection scheme for experiments and how the
optimal detection can change depending on the pulse delay time. %
Finally, when all pulses are well-separated ($\tau_{21}=-\tau_{22^\prime}=2.4\Delta_t$, upper left,
cyan), each resonance condition is independent and both $E_1$ and $E_2$ require FID buildup to
produce final output. %
The resulting line shape is narrow in all directions. %
Again, the emitted frequency does not depend on $\omega_1$, yet the monochromator resolves the
final coherence at frequency $\omega_1$. %
Since the driven part of the final interaction comes from $E_{2^\prime}$, and since the
monochromator track $\omega_1$, the output signal will increase when
$\omega_1=\omega_{2^\prime}$. %
As a result, the line shape acquires a diagonal character. %
The changes in line shape seen in \autoref{mix:fig:pw1} have significant ramifications for the
interpretations and strategies of MR-CMDS in the mixed domain. %
Time-gating has been used to isolate the 2D spectra of a certain
time-ordering \cite{MeyerKentA2004a, PakoulevAndreiV2006a, DonaldsonPaulMurray2007b}, but here we
show that time-gating itself causes significant line shape changes to the isolated pathways. %
The phenomenon of time-gating can cause frequency and delay axes to become functional of each other
in unexpected ways. %
\subsection{Temporal pathway discrimination} % ---------------------------------------------------
\begin{figure}
\includegraphics[width=\linewidth]{"mixed_domain/delay space ratios"}
\caption[2D delay response for different relative dephasing rates.]{
Comparison of the 2D delay response for different relative dephasing rates (labeled atop each
column).
All pulses are tuned to exact resonance.
In each 2D delay plot, the signal amplitude is depicted by the colors.
The black contour lines show signal purity, $P$ (see \autoref{mix:eqn:P}), with purity values
denoted on each contour.
The small plots above each 2D delay plot examine a $\tau_{22^\prime}$ slice of the delay
response ($\tau_{21}=0$).
The plot shows the total signal (black), as well as the component time-orderings VI (orange), V
(purple), and III or I (teal).
}
\label{mix:fig:delay_purity}
\end{figure}
In the last section we showed how a single pathway's spectra can evolve with delay due to pulse
effects and time gating. %
In real experiments, evolution with delay is further complicated by the six time-orderings/sixteen
pathways present in our three-beam experiment (see \autoref{mix:fig:WMELs}). %
Each time-ordering has different resonance conditions. %
When signal is collected near pulse overlap, multiple time-orderings contribute. %
To identify these effects, we start by considering how strongly time-orderings are isolated at each
delay coordinate. %
While the general idea of using time delays to enhance certain time-ordered regions is widely
applied, quantitation of this discrimination is rarely explored. %
Because the temporal profile of the signal is dependent on both the excitation pulse profile and
the decay dynamics of the coherence itself, quantitation of pathway discrimination requires
simulation. %
\autoref{mix:fig:delay_purity} shows the 2D delay space with all pathways present for
$\omega_1,\omega_2=\omega_{10}$. %
It illustrates the interplay of pulse width and system decay rates on the isolation of time-ordered
pathways. %
The color bar shows the signal amplitude. %
Signal is symmetric about the $\tau_{21}=\tau_{22^\prime}$ line because when $\omega_1=\omega_2$,
$E_1$ and $E_{2^\prime}$ interactions are interchangeable:
$S_\text{tot}(\tau_{21},\tau_{22^\prime})=S_\text{tot}(\tau_{22^\prime}, \tau_{21})$. %
The overlaid black contours represent signal ``purity,'' $P$, defined as the relative amount of
signal that comes from the dominant pathway at that delay value:
\begin{equation} \label{mix:eqn:P}
P(\tau_{21},\tau_{22^\prime})=\frac{\max \left\{S_L\left( \tau_{21},\tau_{22^\prime} \right)\right\}}
{\sum_L S_L\left( \tau_{21},\tau_{22^\prime} \right)}.
\end{equation}
The dominant pathway ($\max{\left\{ S_L \left( \tau_{21},\tau_{22^\prime} \right) \right\}}$) at
given delays can be inferred by the time-ordered region defined in \autoref{mix:fig:overview}d. %
The contours of purity generally run parallel to the time-ordering boundaries with the exception of
time-ordered regions II and IV, which involve the double quantum coherences that have been
neglected. %
A commonly-employed metric for temporal selectivity is how definitively the pulses are ordered. %
This metric agrees with our simulations. %
The purity contours have a weak dependence on $\Delta_t \Gamma_{10}$ for
$\left|\tau_{22^\prime}\right|/\Delta_t < 1$ or $\left|\tau_{21}\right|/\Delta_t < 1$ where there
is significant pathway overlap and a stronger dependence at larger values where the pathways are
well-isolated. %
Because responses decay exponentially, while pulses decay as Gaussians, there always exist delays
where temporal discrimination is possible. %
As $\Delta_t\Gamma_{10}\rightarrow \infty$, however, such discrimination is only achieved at
vanishing signal intensities; the contour of $P=0.99$ across our systems highlights this trend. %
\subsection{Multidimensional line shape dependence on pulse delay time}
\begin{figure}
\includegraphics[width=\linewidth]{"mixed_domain/spectral evolution"}
\caption[Evolution of the 2D frequency response.]{
Evolution of the 2D frequency response as a function of $\tau_{21}$ (labeled inset) and the
influence of the relative dephasing rate ($\Gamma_{10}\Delta_t=0.5$ (red), $1.0$ (green), and
$2.0$ (blue)).
In all plots $\tau_{22^\prime}=0$. To ease comparison between different dephasing rates, the
colored line contours (showing the half-maximum) for all three relative dephasing rates are
overlaid.
The colored histograms below each 2D frequency plot show the relative weights of each
time-ordering for each relative dephasing rate.
Contributions from V and VI are grouped together because they have equal weights at
$\tau_{22^\prime}=0$.
}
\label{mix:fig:hom_2d_spectra}
\end{figure}
In the previous sections we showed how pathway spectra and weights evolve with delay. %
This section ties the two concepts together by exploring the evolution of the spectral line shape
over a span of $\tau_{21}$ delay times that include all pathways. %
It is a common practice to explore spectral evolution against $\tau_{21}$ because this delay axis
shows population evolution in a manner analogous to pump-probe spectroscopies. %
The $\vec{k}_2$ and $\vec{k}_{2^\prime}$ interactions correspond to the pump, and the $\vec{k}_1$
interaction corresponds to the probe. %
Time-orderings V and VI are the normal pump-probe time-orderings, time-ordering III is a mixed
pump-probe-pump ordering (so-called pump polarization coupling), and time-ordering I is the
probe-pump ordering (so-called perturbed FID). %
Scanning $\tau_{21}$ through pulse overlap complicates interpretation of the line shape due to the
changing nature and balance of the contributing time-orderings. %
At $\tau_{21}>0$, time-ordering I dominates; at $\tau_{21}=0$, all time-orderings contribute
equally; at $\tau_{21}<0$ time-orderings V and VI dominate (\autoref{mix:fig:delay_purity}). %
Conventional pump-probe techniques recognized these complications long ago, \cite{BritoCruzCH1988a,
PalfreySL1985a} but the extension of these effects to MR-CMDS has not previously been done. %
\autoref{mix:fig:hom_2d_spectra} shows the MR-CMDS spectra, as well as histograms of the pathway
weights, while scanning $\tau_{21}$ through pulse overlap. %
The colored histogram bars and line shape contours correspond to different values of the relative
dephasing rate, $\Gamma_{10}\Delta_t$. %
The contour is the half-maximum of the line shape. %
The dependence of the line shape amplitude on $\tau_{21}$ can be inferred from
\autoref{mix:fig:delay_purity}. %
The qualitative trend, as $\tau_{21}$ goes from positive to negative delays, is a change from
diagonal/compressed line shapes to much broader resonances with no correlation ($\omega_1$ and
$\omega_2$ interact with independent resonances). %
Such spectral changes could be misinterpreted as spectral diffusion, where the line shape changes
from correlated to uncorrelated as population time increases due to system dynamics. %
The system dynamics included here, however, contain no structure that would allow for such
diffusion. %
Rather, the spectral changes reflect the changes in the majority pathway contribution, starting
with time-ordering I pathways, proceeding to an equal admixture of I, III, V, and VI, and finishing
at an equal balance of V and VI when $E_1$ arrives well after $E_2$ and $E_{2^\prime}$. %
Time-orderings I and III both exhibit a spectral correlation in $\omega_1$ and $\omega_2$ when
driven, but time-orderings V and VI do not. %
Moreover, such spectral correlation is forced near zero delay because the pulses time-gate the
driven signals of the first two induced polarizations. %
The monochromator detection also plays a dynamic role, because time-orderings V always VI always
emit a signal at the monochromator frequency, while in time-orderings I and III the emitted
frequency is not defined by $\omega_1$, as discussed above. %
When we isolate time-orderings V and VI, we can maintain the proper scaling of FID bandwidth in the
$\omega_1$ direction because our monochromator can gate the final coherence. %
This gating is not possible in time-orderings I and III because the final coherence frequency is
determined by $\omega_{2^\prime}$ which is identical to $\omega_2$. %
There are differences in the line shapes for the different values of the relative dephasing rate,
$\Gamma_{10}\Delta_t$. %
The spectral correlation at $\tau_{21}/\Delta_t=2$ decreases in strength as $\Gamma_{10}\Delta_t$
decreases. %
As we illustrated in \autoref{mix:fig:pw1}, this spectral correlation is a signature of driven signal
from temporal overlap of $E_1$ and $E_2$; the loss of spectral correlation reflects the increased
prominence of FID in the first coherence as the field-matter interactions become more impulsive. %
This increased prominence of FID also reflects an increase in signal strength, as shown by
$\tau_{21}$ traces in \autoref{mix:fig:delay_purity}. %
When all pulses are completely overlapped, ($\tau_{21}=0$), each of the line shapes exhibit
spectral correlation. %
At $\tau_{21}/\Delta_t=-2$, the line shape shrinks as $\Gamma_{10}\Delta_t$ decreases, with the
elongation direction changing from horizontal to vertical. %
The general shrinking reflects the narrowing homogeneous linewidth of the $\omega_{10}$
resonance. %
In all cases, the horizontal line shape corresponds to the homogeneous linewidth because the narrow
bandpass monochromator resolves the final $\omega_1$ resonance. %
The change in elongation direction is due to the resolving power of $\omega_2$. %
At $\Gamma_{10}\Delta_t=0.5$, the resonance is broader than our pulse bandwidth and is fully
resolved vertically. %
It is narrower than the $\omega_1$ resonance because time-orderings V and VI interfere to isolate
only the absorptive line shape along $\omega_2$. %
This narrowing, however, is unresolvable when the pulse bandwidth becomes broader than that of the
resonance, which gives rise to a vertically elongated signal when $\Gamma_{10}\Delta_t=0.5$. %
\begin{figure}
\includegraphics[width=\linewidth]{"mixed_domain/wigners"}
\caption[Wigners.]{
Transient ($\omega_1$) line shapes and their dependence on $\omega_2$ frequency.
The relative dephasing rate is $\Gamma_{10}\Delta_t=1$ and $\tau_{22^\prime}=0$.
For each plot, the corresponding $\omega_2$ value is shown as a light gray vertical line.}
\label{mix:fig:wigners}
\end{figure}
It is also common to represent data as ``Wigner plots,'' where one axis is delay and the other is
frequency. \cite{KohlerDanielDavid2014a, AubockGerald2012a, CzechKyleJonathan2015a,
PakoulevAndreiV2007a} %
In \autoref{mix:fig:wigners} we show five $\tau_{21},\omega_1$ plots for varying $\omega_2$ with
$\tau_{22^\prime}=0$. %
The plots are the analogue to the most common multidimensional experiment of Transient Absorption
spectroscopy, where the non-linear probe spectrum is plotted as a function of the pump-probe
delay. %
For each plot, the $\omega_2$ frequency is denoted by a vertical gray line. %
Each Wigner plot is scaled to its own dynamic range to emphasize the dependence on $\omega_2$. %
The dramatic line shape changes between positive and negative delays can be seen. %
This representation also highlights the asymmetric broadening of the $\omega_1$ line shape near
pulse overlap when $\omega_2$ becomes non-resonant. %
Again, these features can resemble spectral diffusion even though our system is homogeneous. %
\subsection{Inhomogeneous broadening} \label{mix:sec:res_inhom} % --------------------------------
\begin{figure}
\includegraphics[width=0.5\linewidth]{"mixed_domain/inhom delay space ratios"}
\caption[2D delay response with inhomogeneity.]{
2D delay response for $\Gamma_{10}\Delta_t=1$ with sample inhomogeneity. %
All pulses are tuned to exact resonance. %
The colors depict the signal amplitude. %
The black contour lines show signal purity, $P$ (see \autoref{mix:eqn:P}), with purity values
denoted on each contour. %
The thick yellow line denotes the peak amplitude position that is used for 3PEPS analysis. %
The small plot above each 2D delay plot examines a $\tau_{22^\prime}$ slice of the delay
response ($\tau_{21}=0$). %
The plot shows the total signal (black), as well as the component time-orderings VI (orange), V
(purple), III (teal, dashed), and I (teal, solid). %
}
\label{mix:fig:delay_inhom}
\end{figure}
With the homogeneous system characterized, we can now consider the effect of inhomogeneity. %
For inhomogeneous systems, time-orderings III and V are enhanced because their final coherence will
rephase to form a photon echo, whereas time-orderings I and VI will not. %
In delay space, this rephasing appears as a shift of signal to time-ordered regions III and V that
persists for all population times. %
\autoref{mix:fig:delay_inhom} shows the calculated spectra for relative dephasing rate
$\Gamma_{10}\Delta_t=1$ with a frequency broadening function of width
$\Delta_{\text{inhom}}=0.441\Gamma_{10}$. %
The inhomogeneity makes it easier to temporally isolate the rephasing pathways and harder to
isolate the non-rephasing pathways, as shown by the purity contours. %
A common metric of rephasing in delay space is the 3PEPS
measurement. \cite{WeinerAM1985a, FlemingGrahmR1998a, DeBoeijWimP1998a, SalvadorMayroseR2008a} %
In 3PEPS, one measures the signal as the first coherence time, $\tau$, is scanned across both
rephasing and non-rephasing pathways while keeping population time, $T$, constant. %
The position of the peak is measured; a peak shifted away from $\tau=0$ reflects the rephasing
ability of the system. %
An inhomogeneous system will emit a photon echo in the rephasing pathway, enhancing signal in the
rephasing time-ordering and creating the peak shift. %
In our 2D delay space, the $\tau$ trace can be defined if we assume $E_2$ and $E_{2^\prime}$ create
the population (time-orderings V and VI). %
The trace runs parallel to the III-V time-ordering boundary (diagonal) if $\tau_{22^\prime}<0$ and
runs parallel to the IV-VI time-ordering boundary (horizontal) if $\tau_{22^\prime}>0$, and both
intersect at $\tau_{22^\prime}=0$; the $-\tau_{21}$ value at this intersection is $T$. %
In our 2D delay plots (\autoref{mix:fig:delay_purity}, \autoref{mix:fig:delay_inhom}), the peak shift is seen as the diagonal displacement of the signal peak from the $\tau_{21}=0$ vertical line. %
\autoref{mix:fig:delay_inhom} highlights the peak shift profile as a function of population time with
the yellow trace; it is easily verified that our static inhomogneous system exhibits a non-zero
peak shift value for all population times. %
The unanticipated feature of the 3PEPS analysis is the dependence on $T$. %
Even though our inhomogeneity is static, the peak shift is maximal at $T=0$ and dissipates as $T$
increases, mimicking spectral diffusion. %
This dynamic arises from signal overlap with time-ordering III, which uses $E_2$ and $E_1$ as the
first two interactions ,and merely reflects $E_1$ and $E_2$ temporal overlap. %
At $T=0$, the $\tau$ trace gives two ways to make a rephasing pathway (time-orderings III and V)
and only one way to make a non-rephasing pathway (time-ordering VI). %
This pathway asymmetry shifts signal away from $\tau=0$ into the rephasing direction. %
At large $T$ (large $\tau_{21}$), time-ordering III is not viable and pathway asymmetry
disappears. %
Peak shifts imply inhomogeneity only when time-orderings V and III are minimally contaminated by
each other i.e. at population times that exceed pulse overlap. %
This fact is easily illustrated by the dynamics of homogeneous system (Fig.
\autoref{mix:fig:delay_purity}); even though the homogeneous system cannot rephase, there is a
non-zero peak shift near $T=0$. %
The contamination of the 3PEPS measurement at pulse overlap is well-known and is described in some
studies, \cite{DeBoeijWimP1996a, AgarwalRitesh2002a} but the dependence of pulse and system properties on the
distortion has not been investigated previously. %
Peak-shifting due to pulse overlap is less important when $\omega_1\neq\omega_2$ because
time-ordering III is decoupled by detuning. %
\begin{figure}
\includegraphics[width=\linewidth]{"mixed_domain/inhom spectral evolution"}
\caption[Spectral evolution of an inhomogenious system.]{
Same as \autoref{mix:fig:hom_2d_spectra}, but each system has inhomogeneity
($\Delta_{\text{inhom}}=0.441\Gamma_{10}$).
Relative dephasing rates are $\Gamma_{10}\Delta_t=0.5$ (red), $1.0$ (green), and $2.0$ (blue).
In all plots $\tau_{22^\prime}=0$.
To ease comparison between different dephasing rates, the colored line shapes of all three
systems are overlaid.
Each 2D plot shows a single representative contour (half-maximum) for each
$\Gamma_{10}\Delta_t$ value.
The colored histograms below each 2D frequency plot show the relative weights of each
time-ordering for each 2D frequency plot.
In contrast to \autoref{mix:fig:hom_2d_spectra}, inhomogeneity makes the relative contributions of
time-orderings V and VI unequal.
}
\label{mix:fig:inhom_2d_spectra}
\end{figure}
In frequency space, spectral elongation along the diagonal is the signature of inhomogeneous
broadening. %
\autoref{mix:fig:inhom_2d_spectra} shows the line shape changes of a Gaussian inhomogeneous
distribution. %
All systems are broadened by a distribution proportional to their dephasing bandwidth. %
As expected, the sequence again shows a gradual broadening along the $\omega_1$ axis, with a strong
spectral correlation at early delays ($\tau_{21}>0$) for the more driven signals. %
The anti-diagonal width at early delays (e.g. \autoref{mix:fig:inhom_2d_spectra},
$\tau_{21}=2.0\Delta_t$) again depends on the pulse bandwidth and the monochromator slit width. %
At delay values that isolate time-orderings V and VI, however, the line shapes retain diagonal
character, showing the characteristic balance of homogeneous and inhomogeneous width. %
\section{Discussion} % ---------------------------------------------------------------------------
\subsection{An intuitive picture of pulse effects}
Our chosen values of the relative dephasing time, $\Gamma_{10}\Delta_t$, describe experiments where
neither the impulsive nor driven limit unilaterally applies. %
We have illustrated that in this intermediate regime, the multidimensional spectra contain
attributes of both limits, and that it is possible to judge when these attributes apply. %
In our three-pulse experiment the second and third pulses time-gate coherences and populations
produced by the previous pulse(s), and the monochromator frequency-gates the final coherence. %
Time-gating isolates different properties of the coherences and populations. Consequently, spectra
evolve against delay. %
For any delay coordinate, one can develop qualitative line shape expectations by considering the following three principles:
\begin{enumerate}
\item When time-gating during the pulse, the system pins to the driving frequency with a buildup efficiency determined by resonance.
\item When time-gating after the pulse, the FID dominates the system response.
\item The emitted signal field contains both FID and driven components; the $\omega_{\text{out}} = \omega_1$ component is isolated by the tracking monochromator.
\end{enumerate}
\autoref{mix:fig:fid_dpr} illustrates principles 1 and 2 and \autoref{mix:fig:fid_detuning}
illustrates principle 2 and 3. %
\autoref{mix:fig:pw1} provides a detailed example of the relationship between these principles and the
multidimensional line shape changes for different delay times. %
The principles presented above apply to a single pathway. %
For rapidly dephasing systems it is difficult to achieve complete pathway discrimination, as shown
in \autoref{mix:fig:delay_purity}. %
In such situations the interference between pathways must be considered to predict the line
shape. %
The relative weight of each pathway to the interference can be approximated by the extent of pulse
overlap. %
The pathway weights exchange when scanning across pulse overlap, which creates the dramatic line
shape changes observed in Figures \ref{mix:fig:hom_2d_spectra} and
\ref{mix:fig:inhom_2d_spectra}. %
\subsection{Conditional validity of the driven limit}
We have shown that the driven limit misses details of the line shape if $\Gamma_{10} \Delta_t
\approx 1$, but we have also reasoned that in certain conditions the driven limit can approximate
the response well (see principle 1). %
Here we examine the line shape at delay values that demonstrate this agreement. %
Fig. \ref{mix:fig:steady_state} compares the results of our numerical simulation (third column) with
the driven limit expressions for populations where $\Gamma_{11}\Delta_t=0$ (first column) or $1$
(second column). %
The top and bottom rows compare the line shapes when $\left(\tau_{22^\prime},
\tau_{21}=(0,0)\right)$ and $(0,-4\Delta_t)$, respectively. %
The third column demonstrates the agreement between the driven limit approximations with the
simulation by comparing the diagonal and anti-diagonal cross-sections of the 2D spectra. %
% TODO: [ ] population resonance is not clear
Note the very sharp diagonal feature that appears for $(\tau_{21},\tau_{22^\prime}) = (0,0)$ and $
\Gamma_{11}=0$; this is due to population resonance in time-orderings I and III. %
This expression is inaccurate: the narrow resonance is only observed when pulse durations are much
longer than the coherence time. %
A comparison of picosecond and femtosecond studies of quantum dot exciton line shapes
(\textcite{YursLenaA2011a} and \textcite{KohlerDanielDavid2014a}, respectively) demonstrates this
difference well. %
The driven equation fails to reproduce our numerical simulations here because resonant excitation
of the population is impulsive; the experiment time-gates only the rise time of the population, yet
driven theory predicts the resonance to be vanishingly narrow ($\Gamma_{11}=0$). %
In light of this, one can approximate this time-gating effect by substituting population lifetime
with the pulse duration ($\Gamma_{11}\Delta_t=1$), which gives good agreement with the numerical
simulation (third column). %
When $\tau_{22^\prime}=0$ and $\tau_{21}<\Delta_t$, signals can also be approximated by driven
signal (\autoref{mix:fig:steady_state} bottom row). %
Only time-orderings V and VI are relevant. %
The intermediate population resonance is still impulsive but it depends on
$\omega_{2^\prime}-\omega_2$ which is not explored in our 2D frequency space. %
\begin{figure}
\includegraphics[width=\linewidth]{"mixed_domain/steady state"}
\caption[Conditional validity of the driven limit.]{
Comparing approximate expressions of the 2D frequency response with the directly integrated
response. %
$\Gamma_{10}\Delta_t=1$. %
The top row compares the 2D response of all time-orderings ($\tau_{21}=0$) and the bottom row
compares the response of time-orderings V and VI ($\tau_{21}=-4\Delta_t$). %
First column: The driven limit response. Note the narrow diagonal resonance for $\tau_{21}=0$.
Second column: Same as the first column, but with ad hoc substitution $\Gamma_{11}=\Delta_t$.
Third column: The directly integrated response. %
}
\label{mix:fig:steady_state}
\end{figure}
\subsection{Extracting true material correlation} % ----------------------------------------------
\begin{figure}
\includegraphics[width=0.5\linewidth]{"mixed_domain/metrics"}
\caption[Metrics of correlation.]{
Temporal (3PEPS) and spectral (ellipticity) metrics of correlation and their relation to the
true system inhomogeneity. %
The left column plots the relationship at pulse overlap ($T=0$) and the right column plots the
relationship at a delay where driven correlations are removed ($T=4\Delta_t$). %
For the ellipticity measurements, $\tau_{22^\prime}=0$. %
In each case, the two metrics are plotted directly against system inhomogeneity (top and middle
row) and against each other (bottom row). %
Colored lines guide the eyes for systems with equal relative dephasing rates
($\Gamma_{10}\Delta_t$, see upper legend), while the area of the data point marker indicates
the relative inhomogeneity ($\Delta_{\text{inhom}}/\Gamma_{10}$, see lower legend). %
Gray lines indicate contours of constant relative inhomogeneity (scatter points with the same
area are connected). %
}
\label{mix:fig:metrics}
\end{figure}
We have shown that pulse effects mimic the qualitative signatures of inhomogeneity. %
Here we address how one can extract true system inhomogeneity in light of these effects. %
We focus on two ubiquitous metrics of inhomogeneity: 3PEPS for the time domain and
ellipticity for the frequency domain \cite{KwacKijeong2003a, OkumuraKo1999a}. %
There are many ways to characterize the ellipticity of a peak shape. %
We adopt the convention $\mathcal{E} = \left(a^2-b^2\right) / \left(a^2+b^2\right)$, where $a$ is
the diagonal width and $b$ is the antidiagonal width. %
In the driven (impulsive) limit, ellipticity (3PEPS) corresponds to the frequency correlation
function and uniquely extracts the inhomogeneity of the models presented here. %
In their respective limits, the metrics give values proportional to the inhomogeneity. %
\autoref{mix:fig:metrics} shows the results of this characterization for all $\Delta_\text{inhom}$ and
$\Gamma_{10}\Delta_t$ values explored in this work. %
We study how the correlations between the two metrics depend on the relative dephasing rate,
$\Gamma_{10}\Delta_t$, the absolute inhomogeneity, $\Delta_\text{inhom} / \Delta_\omega$, the
relative inhomogeneity $\Delta_\text{inhom} / \Gamma_{10}$, and the population time delay. %
The top row shows the correlations of the $\Delta_\text{3PEPS} / \Delta_t$ 3PEPS metric that
represents the normalized coherence delay time required to reach the peak intensity. %
The upper right graph shows the correlations for a population time delay of $T = 4\Delta_t$ that
isolates the V and VI time-orderings. %
For this time delay, the $\Delta_\text{3PEPS} / \Delta_t$ metric works well for all dephasing times
of $\Gamma_{10}\Delta_t$ when the relative inhomogeneity is $\Delta_\text{inhom} / \Delta_\omega
\ll 1$. %
It becomes independent of $\Delta_\text{inhom} / \Delta_\omega$ when $\Delta_\text{inhom} /
\Delta_\omega > 1$. %
This saturation results because the frequency bandwidth of the excitation pulses becomes smaller
than the inhomogeneous width and only a portion of the inhomogeneous ensemble contributes to the
3PEPS experiment. \cite{WeinerAM1985a} %
The corresponding graph for $T = 0$ shows a large peak shift occurs, even without inhomogeneity.
In this case, the peak shift depends on pathway overlap, as discussed in
\autoref{mix:sec:res_inhom}. %
The middle row in \autoref{mix:fig:metrics} shows the ellipticity dependence on the relative dephasing
rate and inhomogeneity assuming the measurement is performed when the first two pulses are
temporally overlapped ($\tau_{22^\prime}=0$). %
For a $T=4\Delta_t$ population time, the ellipticity is proportional to the inhomogeneity until
$\Delta_\text{inhom} / \Delta_\omega \ll 1$ where the excitation bandwidth is wide compared with
the inhomogeneity. %
Unlike 3PEPS, saturation is not observed because pulse bandwidth does not limit the frequency range
scanned. %
The 3PEPS and ellipticity metrics are therefore complementary since 3PEPS works well for
$\Delta_\text{inhom} / \Delta_\omega \ll 1$ and ellipticity works well for $\Delta_\text{inhom} /
\Delta_\omega \gg 1$. %
When all pulses are temporally overlapped at $T = 0$, the ellipticity is only weakly dependent on
the inhomogeneity and dephasing rate. %
The ellipticity is instead dominated by the dependence on the excitation pulse frequency
differences of time-orderings I and III that become important at pulse overlap. %
It is clear from the previous discussion that both metrics depend on the dephasing and
inhomogeneity. %
The dephasing can be measured independently in the frequency or time domain, depending upon whether
the dephasing is very fast or slow, respectively. %
In the mixed frequency/time domain, measurement of the dephasing becomes more difficult. %
One strategy to address this challenge is to use both the 3PEPS and ellipticity metrics. %
The bottom row in \autoref{mix:fig:metrics} plots 3PEPS against ellipticity to show how the
relationship between the metrics changes for different amounts of dephasing and inhomogeneity. %
The anti-diagonal contours of constant relative inhomogeneity show that these metrics are
complementary and can serve to extract the system correlation parameters. %
Importantly, the metrics are uniquely mapped both in the presence and absence of pulse-induced
effects (demonstrated by $T = 0$ and $T = 4\Delta_t$, respectively). %
The combined metrics can be used to determine correlation at $T = 0$, but the correlation-inducing
pulse effects give a mapping significantly different than at $T = 4\Delta_t$. %
At $T = 0$, 3PEPS is almost nonresponsive to inhomogeneity; instead, it is an almost independent
characterization of the pure dephasing. %
In fact, the $T=0$ trace is equivalent to the original photon echo traces used to resolve pure
dephasing rates. \cite{AartsmaThijsJ1976a} %
Both metrics are offset due to the pulse overlap effects. %
Accordingly, the region to the left of homogeneous contour is non-physical, because it represents
observed correlations that are less than that given by pulse overlap effects. %
If the metrics are measured as a function of $T$, the mapping gradually changes from the left
figure to the right figure in accordance with the pulse overlap. %
Both metrics will show a decrease, even with static inhomogeneity. %
If a system has spectral diffusion, the mapping at late times will disagree with the mapping at
early times; both ellipticity and 3PEPS will be smaller at later times than predicted by the change
in mappings alone. %
\section{Conclusion} % ---------------------------------------------------------------------------
This study provides a framework to describe and disentangle the influence of the excitation pulses
in mixed-domain ultrafast spectroscopy. %
We analyzed the features of mixed domain spectroscopy through detailed simulations of MR-CMDS
signals. %
When pulse durations are similar to coherence times, resolution is compromised by time-bandwidth
uncertainty and the complex mixture of driven and FID response. %
The dimensionless quantity $\Delta_t \left(\Gamma_f + i\kappa_f \Omega_{fx}\right)$ captures the
balance of driven and FID character in a single field-matter interaction. %
In the nonlinear experiment, with multiple field-matter interactions, this balance is also
controlled by pulse delays and frequency-resolved detection. %
Our analysis shows how these effects can be intuitive. %
The dynamic nature of pulse effects can lead to misleading changes to spectra when delays are
changed. %
When delays separate pulses, the spectral line shapes of individual pathways qualitatively change
because the delays isolate FID contributions and de-emphasize driven response. %
When delays are scanned across pulse overlap, the weights of individual pathways change, further
changing the line shapes. %
In a real system, these changes would all be present in addition to actual dynamics and spectral
changes of the material. %
Finally, we find that, in either frequency or time domain, pulse effects mimic signatures of
ultrafast inhomogeneity. %
Even homogeneous systems take on these signatures. %
For mixed domain experiments, pulse effects induce spectral ellipticity and photon echo signatures,
even in homogeneous systems. %
Driven character gives rise to pathway overlap peak shifting in the 2D delay response, which
artificially produces rephasing near pulse overlap. %
Driven character also produces resonances that depend on $\omega_1-\omega_2$ near pulse overlap. %
Determination of the homogeneous and inhomogeneous broadening at ultrashort times is only possible
by performing correlation analysis in both the frequency and time domain. %
|