diff options
author | Blaise Thompson <blaise@untzag.com> | 2018-04-15 00:16:26 -0500 |
---|---|---|
committer | Blaise Thompson <blaise@untzag.com> | 2018-04-15 00:16:26 -0500 |
commit | b75d9d8f2ba798fbbadc975a789cf2615b743328 (patch) | |
tree | 404c03de1aae4dc9f05194dc2d4e6e36203dafe0 /procedures | |
parent | 4c38970579d9bd994b18f10ab9fd3956beff4929 (diff) |
2018-04-15 00:16
Diffstat (limited to 'procedures')
-rw-r--r-- | procedures/chapter.tex | 49 |
1 files changed, 23 insertions, 26 deletions
diff --git a/procedures/chapter.tex b/procedures/chapter.tex index fd04bc3..8c8d0b2 100644 --- a/procedures/chapter.tex +++ b/procedures/chapter.tex @@ -1,7 +1,17 @@ \chapter{Procedures} -\clearpage -\section{``Six-month'' maintenance} % ------------------------------------------------------------ +\begin{dquote} + These are the vegetables we must eat before we can have our pixels of ice cream. + + \dsignature{Wright Group saying}. +\end{dquote} + +\vfill + +In this chapter I document the various procedures that I have performed to maintain the MR-CMDS +instruments. % + +\section{``Six-month'' maintenance} % ============================================================ The laser system that the Wright Group's MR-CMDS instruments use requires regular maintenance. % Each component is sensitive to lab conditions such as temperature, humidity, and vibrations. % @@ -13,7 +23,6 @@ Historically, the Wright Group has engaged in reactive maintenance: a ``fix it i don't touch otherwise'' kind of approach. % This approach makes a lot of sense for instruments that are quick to fix, and have few active users. % - I instituted a proactive, regular maintenance procedure (described below) that has improved the predictability of instrumental performance. % Predictability is key for instruments with multiple users. % @@ -48,9 +57,9 @@ Procedure: \end{denumerate} \item Chiller maintenance. \begin{ditemize} - \item See x - \item See y - \item See z + \item See \autoref{prc:sec:lytron} + \item See \autoref{prc:sec:polyscience} + \item See \autoref{prc:sec:neslab} \end{ditemize} \item Lab cleaning. \begin{denumerate} @@ -68,13 +77,9 @@ Procedure: \end{ditemize} \item Let lab sit overnight to allow dust to settle and the air to dehumidify. \item Start up system again. - \begin{denumerate} - \item TODO - \end{denumerate} \end{denumerate} -\clearpage -\section{Lytron Kodiak RC006} % ------------------------------------------------------------------ +\section{Lytron Kodiak RC006} \label{prc:sec:lytron} % =========================================== We have one Lytron Kodiak RC006: Model Number RC006G03BB1C002, Serial Number 739383-02. @@ -107,10 +112,7 @@ These seem to have been fixed by adding a “high” flow loop connecting the ou chiller. % Ideally the pressure drop across this loop is sufficient to still drive fluid through the laser. % -% TODO: figure - -\clearpage -\section{PolyScience 6000 Series} % -------------------------------------------------------------- +\section{PolyScience 6000 Series} \label{prc:sec:polyscience} % ================================== We own two PolyScience chillers---different models but functionally equivalent. @@ -142,8 +144,7 @@ Regular chiller maintenance: \end{denumerate} \end{denumerate} -\clearpage -\section{NesLab Merlin M33} % ==================================================================== +\section{NesLab Merlin M33} \label{prc:sec:neslab} % ============================================= We have one NesLab Merlin M33 Chiller, Serial Number 106227049. % @@ -171,7 +172,6 @@ Regular chiller maintenance % TODO: figures -\clearpage \section{Calibrating the 407A} % ================================================================= Calibrating the 407.A @@ -180,11 +180,11 @@ You may sometimes notice that the zero position changes dramatically from sensit sensitivity with the 407A. If this happens, iterate through the following until zero stays consistent: -Use the fine adjust (knob on side) to zero the 407A on the highest sensitivity - -Use the front adjust (flathead screwdriver needed) to zero on the lowest sensitivity - -\clearpage +\begin{ditemize} + \item Use the fine adjust (knob on side) to zero the 407A on the highest sensitivity. + \item Use the front adjust (flathead screwdriver needed) to zero on the lowest sensitivity. +\end{ditemize} + \section{Millenia} % ============================================================================= \subsection{Startup} % --------------------------------------------------------------------------- @@ -228,7 +228,6 @@ These allow us to record things like Diode hours. % Service mode can be buggy, so it's best to leave the Millenia in normal Power mode during regular operation. % -\clearpage \section{Spitfire Pro} % ========================================================================= Only tune up the Spitfire if you need to, and do not treat it casually---set aside an entire @@ -429,7 +428,6 @@ Since we have not messed with the stretcher frequently this guide cannot be trus square mirrors preceding the compressor. \end{denumerate} -\clearpage \section{TOPAS-C} % ============================================================================== % TODO: figure @@ -668,7 +666,6 @@ Input poynting is adjusted to ensure good alignment through L1 and L2 into D1. \item Measure and record power---should be over 600 mW. \end{denumerate} -\clearpage \section{MicroHR Monochromator} % ================================================================ Visible Grating. % |