From f47e1abac351ed10110c9d982aea76ca73dd30a6 Mon Sep 17 00:00:00 2001 From: Blaise Thompson Date: Wed, 31 Jan 2018 18:37:43 -0600 Subject: 2018-01-31 18:37 --- procedures/chapter.tex | 34 +++++++++++++++++++++++++++++++++- 1 file changed, 33 insertions(+), 1 deletion(-) (limited to 'procedures/chapter.tex') diff --git a/procedures/chapter.tex b/procedures/chapter.tex index 3cb42bf..4ac67ed 100644 --- a/procedures/chapter.tex +++ b/procedures/chapter.tex @@ -1 +1,33 @@ -\chapter{Procedures} \ No newline at end of file +\chapter{Procedures} + +\section{Aligining TOPAS-C} + +\section{Aligning Spitfire PRO} + +\section{Air Handling} + +\section{Six Month Maintenance} + +\section{Tuning MicroHR Monochromator} + +Visible Grating. % + +Align the HeNe as perpendicular as possible to the monochromator entrance slit. % + +Move the grating angle until the HeNe falls on the exit slit. % + +Shine a flashlight through the entrance slit and observe the colour on the exit slit: if white, +then you are at 0-order (0 nm), if red, then you are at 1st order (632.8 nm). % + +Go to 0-order, narrow the slits, and slowly adjust the angle until the HeNe is going through the +exit slit. % + +Go to Jovin Yvon/utilities and find the motor configuration program. % + +In the Gratings tab, select the 1st grating (1200 line density) and hit Calibrate. % + +In theoretical wavelength, enter 0 nm. % + +In experimental wavelength, enter the wavelength you observe from the control program. % + +Hit set. % -- cgit v1.2.3